Table 4.
Diagnostic performances of SQA using conventional and digital controls at the individual level with fully automated and visual analyses
Accuracy (%) | Sensitivity (%) | Specificity (%) | p value | |
---|---|---|---|---|
Fully automated | < 0.01* | |||
SQA using conventional controls | 80 | 78 | 82 | |
SQA using digital controls | 86 | 89 | 82 | |
Visual | = 0.01* | |||
SQA using conventional controls | 82 | 67 | 100 | |
SQA using digital controls | 84 | 85 | 82 |