Table 1.
Exposure dose en face of each bulb source to 12 different locations along the respirator rack, as denoted in Figure 1
Height (cm) | Irradiance (mW/cm2) |
Exposure dose (mJ/cm2) |
||||
---|---|---|---|---|---|---|
1 | 2 | 3 | 1 | 2 | 3 | |
Light source 1 | ||||||
162.0 | 0.955 | 1.068 | 0.979 | 1261 | 1410 | 1292 |
131.5 | 1.236 | 1.406 | 1.292 | 1632 | 1856 | 1705 |
99.5 | 1.349 | 1.526 | 1.380 | 1781 | 2014 | 1822 |
68.5 | 1.120 | 1.182 | 1.046 | 1478 | 1560 | 1381 |
Light source 2 | ||||||
162.0 | 1.012 | 1.085 | 0.941 | 1336 | 1432 | 1242 |
131.5 | 1.284 | 1.393 | 1.216 | 1695 | 1839 | 1605 |
99.5 | 1.208 | 1.311 | 1.128 | 1595 | 1731 | 1489 |
68.5 | 0.873 | 0.908 | 0.784 | 1153 | 1198 | 1035 |
Goal exposure dose (mJ/cm2) = | 1000 | |||||
Time (seconds) = | 1320 |