Table 2. Statistical analysis of Schaeffer's test.
Result | ||
Analysis of results for Schaeffer's test | (%) | |
Sensitivity | 93.6 | |
Specificity | 100 | |
Positive predictive value | 100 | |
Negative predictive value | 50.8 | |
Analysis | ||
Palmaris longus tendon present intraoperatively | Palmaris longus tendon absent intraoperatively | |
Schaeffer's test positive | 453 | 0 |
Schaeffer's test negative | 31 | 32 |
Note: results and analysis from Schaeffer's test to determine if the palmaris longus tendon was present or absent compared with the direct intraoperative visualization of the tendon. Note that approximately half of the time when the tendon was not detected preoperatively, the tendon was found intraoperatively (negative predictive value of 50.8%). Also, when Schaeffer's test was positive, the tendon was always identified intraoperatively (specificity and positive predictive value were 100%).