Table 5.
Summary of selected intracortical microstimulation studies
Citation | Model system | Electrode type | Reporter of efficacy | Current range applied | Frequency (Hz) | Impedance range 1 kHz | Damage? |
---|---|---|---|---|---|---|---|
Murphey and Maunsell (2008) | Macaque | Platinum/iridium, single electrode | Behavioral task (visual) | Up to 50 μA | 200 | 0.2–1.5 MΩ | No |
Murphey and Maunsell (2007) | Macaque | Platinum/iridium, single electrode | Behavioral task (visual) | 3–12 μA | 200 | 0.2–1.5 MΩ | No |
McCreery et al. (2010) | Feline | Activated iridium, 16 microelectrode array | N/A | 10–20 μA | 50 | Not reported | Yes |
McCreery et al. (2002) | Feline | Activated iridium, 16 microelectrode array | Evoked potentials in medulla | 0–32 μA | 50 | 0.2–1.5 MΩ | No |
Graziano (2005) | Macaque | Tungsten, single microelectrode | Arm movements | 5–100 μA | 100–250 | 1–2 MΩ | No |
Graziano et al. (2002) | Macaque | Tungsten, single microelectrode | Arm movements | Up to 150 μA | 200 | 0.5–2 MΩ | No |
Tehovnik et al. (2003) | Macaque | Platinum/iridium, single microelectrodes | Behavioral task (visual) | Up to 30 μA | 200 | 1–2 MΩ | No |
Tehovnik et al. (2002) | Macaque | Platinum/iridium, single microelectrodes | Behavioral task (visual) | Up to 40 μA | 200 | 1–1.5 MΩ | No |
Romo et al. (2000) | Macaque | Platinum/tungsten, seven microelectrodes | Behavioral task (flutter discrimination) | 65–100 μA | 50–200 | 1–1.5 MΩ | No |
Romo et al. (1998) | Macaque | Platinum/tungsten, seven microelectrodes | Behavioral task (flutter discrimination) | 65–100 μA | 50–200 | 1–1.5 MΩ | No |
Rousche and Normann (1999) | Feline | Platinum, 96 electrode Utah Electrode Array | Behavioral task (auditory) | 100 μA | 25–2000 | 30–149 kΩ | No |
Torab et al. (2011) | Macaque | SIROF, 96 electrode Utah Electrode Array | Behavioral task (visual) | 0–96 μA | 200 | 40 kΩ−2 MΩ | No |
Bradley (2004) | Macaque | Iridium, 192 electrode array | Behavioral task (visual) | 12–20 μA | 200 | 80 kΩ−1.6 MΩ | No |
Bak et al. (1990) | Human | Iridium, arrays of 1–3 electrodes | Verbal report (visual percepts) | 20 μA-2 μA | 100 | Not reported | No |
Schmidt et al. (1996) | Human | Iridium, 12 single electrodes, 13 paired electrodes | Verbal report (visual percepts) | 1–80 μA | 200 | Not reported | No |
The stimulus amplitudes and parameters applied in this study were similar to those used in other microstimulation experiments which evoked sensory percepts. Shown are sample studies for comparison.