Skip to main content
. 2020 Dec 3;20:293. doi: 10.1186/s12874-020-01179-5

Table 5.

Standards for design requirements of studies on reliability or measurement error

Design requirements very good adequate doubtful inadequate NA
1

Were patients stable in the time between the repeated measurements on the construct to be measured?

Relevance: 39/40 (98%) (R2a); wording: 33/40 (83%) (R2)

Yes (evidence provided) Reasons to assume standard was met Unclear No (evidence provided) NA
2

Was the time interval between the repeated measurements appropriate?

Relevance: 40/41 (98%)(R2); wording: 37/41 (90%)(R2)

Yes

Doubtful ,

OR time interval not stated

No NA
3

Were the measurement condition similar for the repeated measurements – except for the condition being evaluated as a source of variation?

Relevance: 37/41 (90%)(R2); wording: 34/41 (83%)(R2)

Yes (evidence provided)

Reasons to assume standard was met,

OR change was unavoidable

Unclear No (evidence provided) NA
4

Did the professional(s) administer the measurement without knowledge of scores or values of other repeated measurement(s) in the same patients?

Relevance: 38/41 (93%)(R2); wording: 27/30 (90%)(R3b)

Yes (evidence provided) Reasons to assume standard was met Unclear No (evidence provided) NA
5

Did the professional(s) assign the scores or determined the values without knowledge of the scores or values of other repeated measurement(s) in the same patients?

Relevance: 38/41 (93%)(R2); wording: 27/30 (90%)(R3)

Yes (evidence provided) Reasons to assume standard was met Unclear No (evidence provided)
6 Were there any other important flaws in the design or statistical methods of the study? c No Minor methodological flaws Yes

a R2: consensus reached in round 2; b R3: consensus reached in round 3; c Standard 6 and the responses of the four-point rating system were not discussed in the Delphi study