Table 2.

Performance Comparison of Each Probe within a Range of 73-97% of Arterial Oxygen Saturation (SaO2)

Probes 1 2 3 4 5 6 7
Arms (%) 2.83 3.98 3.75 6.84 3.43 5.17 3.84
Mean bias (%) -1.10 3.29 3.04 6.01 2.63 4.69 2.70
Precision (%) 2.22 2.25 2.19 2.56 2.16 2.07 2.74

1: WA-100 reusable finger probe, 2: MDNA disposable finger probe, 3: IS-1011 disposable finger probe, 4: CJ340NA disposable finger probe, 5: NellcorTM OxiMax DS-100A reusable finger probe, 6: NellcorTM OxiMax MAX-N disposable finger probe, 7: OXI-PRO DA disposable finger probe. Arms, bias, and precision were calculated by the following equations: Arms=i=1n(SpO2i-SRi)2n, Bias=SpO2i-SRi, Precision=i=1n(SpO2i-SpO2fit,i)2n-2 , where SpO2ij is the ith measured peripheral oxygen saturation (SpO2) value, SRi is the ith measured standard reference value, and SpO2fit,i is the value of the curve fitted to the test data at SRi.