Skip to main content
. 2020 Dec 2;2(12):e0292. doi: 10.1097/CCE.0000000000000292

TABLE 2.

ISO Tests With Performance Outcomes With Drager Apollo Compared With Portsmouth Ventilator

Peak Pressure (cm H2O) Plateau Pressure (cm H2O) Tidal Volume (mL) % Difference
ISO Trial Number Drager PV Drager PV Drager PV Drager PV
1 19.8 21.4 15.2 17.2 482 535 –3.60 7.00
2 33.8 46 27 28.2 473 522 –5.40 4.40
3 32.2 28.9 21.9 25.9 490 497 –2.00 –0.60
4 39.9 47.5 35.7 38.5 461 477 –7.80 –4.60
5 30.5 38.1 22.3 26.1 285 325 –5.00 8.33
6 40.7 61.2 31.8 37.5 275 307 –8.33 2.33
7 40.7 63.9 36 45.5 246 275 –18.00 –8.33

PV = Portsmouth Ventilator.