. 2020 Nov 20;14:590900. doi: 10.3389/fnins.2020.590900

Table 1.

Main parameters of the SPEN and EPI acquisitions and processing used in the various scans presented in this study.

Parameters Full brain SE-EPI Full brain SPEN Cerebellum SPEN Olfactory bulb SPEN (3D) Olfactory bulb development SPEN (2D)
Repetition time (ms) 1,000 1,000 1,000 1,000 1,000
Echo Time (ms) 29.47 39.57 37.30 36.15 35.18
Acquisition time 1 h 57 min 1 h 57 min 2 h 2 min 1 h 59 min 1 h 10 min
Interleaved segments 10 5 2 2 5
Averaged scans 1 2 4 4 16
Double sampling Yes No No No No
Field of View (mm) 15*20*5.4$ 15*17*5.4 8*5*4 8*5*5 15*13*0.3
Matrix size 200*225*54 200*200*54 130*68*54 120*64*64 256*200*2
Regridded readout points 146 155 105 95 195
Digital resolution (μm) 103*89*100 97*85*100 76*74*74 84*78*78 77*65*300
Readout bandwidth (kHz) 500 500 250 250 500
2nd dimension (Phase, SPEN) bandwidth—kHz 12.5 13.5 3.96 5.86 11.7
SPEN's time-bandwidth product N/A 120 35 45 120
Chirp duration (ms) N/A 8.88 8.84 7.68 10.24
Diffusion directions 12 12 16 13 23
b-values (s/mm2) 700 700 700 700 950
Diffusion gradient strength (mT/m) 666 666 666 666 764
Diffusion gradient duration (ms) 2 2 2 2 2
Intergradient delay (ms) 7 7 7 7 7.2
Slice selection gradient (mT/m) 8.6 8.6 9.4 11.7 156.6
FWHH (localized shimming, in Hz, ±10 Hz #) 45 45 70 45 50
Denoising kernel 3*3 8*8 2*2 2*2 2*2
Average SNR of b0 image (±10%#) 45 21 32 25 21
$

EPI's FOV along the blipped dimension had to be increased vis-à-vis SPEN's to avoid folding; the number of acquisition points was increased accordingly to keep equivalent spatial resolutions.

#

Errors correspond to standard deviations through a series of measurements.