. 2020 Dec 1;9(1):2547–2549. doi: 10.1080/22221751.2020.1852893

Table 1.

Performance of NG-Test® and the Wantai ELISA immunoasay for SARS-CoV-2 antibody detection in Healthcare Workers (HCWs), Rennes University Hospital, France 2020.

    IgG  IgM IgG or IgM
LFIA NG-Test®(vs RT-PCR)
  TP 42 42 42
  FN 9 9 9
  Sensitivity 82.5% [71.9%–92.3%] 82.5% [71.9%–92.3%] 82.5% [71.9%–92.3%]
  TN 58 58 58
  FP 1 1 1
  Specificity 98.3% [95.0%–100.0%] 98.3% [95.0%–100.0%] 98.3% [95.0%–100.0%]
Elisa Wantai (vs RT-PCR)
  TP 49 18 49
  FN 2 33 2
  Sensitivity 96.1% [90.8%–100%] 35.3% [22.2%–48.4%] 96.1% [90.8%–100%]
  TN 59 59 59
  FP 0 0 0
  Specificity 100% [93.4%–100%] 100% [93.4%–100%] 100% [93.4%–100%]

TP: true positive, FN: false negative, TN: true negative, FP: false positive.