Table 1.
Performance of NG-Test® and the Wantai ELISA immunoasay for SARS-CoV-2 antibody detection in Healthcare Workers (HCWs), Rennes University Hospital, France 2020.
IgG | IgM | IgG or IgM | ||
---|---|---|---|---|
LFIA NG-Test®(vs RT-PCR) | ||||
TP | 42 | 42 | 42 | |
FN | 9 | 9 | 9 | |
Sensitivity | 82.5% [71.9%–92.3%] | 82.5% [71.9%–92.3%] | 82.5% [71.9%–92.3%] | |
TN | 58 | 58 | 58 | |
FP | 1 | 1 | 1 | |
Specificity | 98.3% [95.0%–100.0%] | 98.3% [95.0%–100.0%] | 98.3% [95.0%–100.0%] | |
Elisa Wantai (vs RT-PCR) | ||||
TP | 49 | 18 | 49 | |
FN | 2 | 33 | 2 | |
Sensitivity | 96.1% [90.8%–100%] | 35.3% [22.2%–48.4%] | 96.1% [90.8%–100%] | |
TN | 59 | 59 | 59 | |
FP | 0 | 0 | 0 | |
Specificity | 100% [93.4%–100%] | 100% [93.4%–100%] | 100% [93.4%–100%] |
TP: true positive, FN: false negative, TN: true negative, FP: false positive.