NPV for Initial Negative Test Results by Assay or Specimen Type

Comparison by Group Initial negative (n = 811) True negative (n = 795) False negative (n = 16) NPV, % (95% CI)
Assay CDC RT-PCR 168 166 2 98.8 (95.7–99.7)
Simplexa Direct 518 506 12 97.7 (96.2–98.6)
TaqPath RT-PCR 125 123 2 98.4 (94.6–99.5)
Specimen type Initial URT 785 769 16 98.0 (96.9–98.7)
Initial LRT 26 26 0 100

LRT, lower respiratory tract; NPV, negative predictive value; URT, upper respiratory tract.