Skip to main content
. Author manuscript; available in PMC: 2021 Dec 1.
Published in final edited form as: Magn Reson Med. 2020 Jun 7;84(6):3054–3070. doi: 10.1002/mrm.28338

FIGURE 6.

FIGURE 6

Scatterplots of the quantitative scores vs the average radiologists’ score based on ranking (1 is best) for the top 4 submissions in all three submission tracks. NMSE, PSNR, and SSIM scores are normalized so that the best score corresponds to a value of 1 for convenient visualization. For multi-coil R=8, the highest ranked submission was also the one that had the highest SSIM, NMSE and PSNR values. For single-coil R=4, only SSIM showed a similar trend as the radiologists scores, while the other two metrics showed almost opposite trends. For the multi-coil R=4 track, the top 4 submissions were very close together with all metrics. Therefore, the results are less conclusive.