Table 2.
Probe | Fiber | Depth of | Axial | Frame | |||
---|---|---|---|---|---|---|---|
diameter | characteristics | penetration | resolution | rate | Limitation | ||
NIRF-OCT | Hara et al. (57) | 2.4 Fr | • OCT: 1320 nm • NIRF: 750 nm | NIRF: 3 mm | OCT: 7 μm | 25.4/s | Slower image acquisition speed |
Zaheer et al. (60) | 2.6 Fr | • OCT: 1,290 nm • NIRF: 749–790 nm | OCT: 1–2 mm | OCT: 7 μm | 100/s | – | |
NIRF-IVUS | Shi et al. (63) | 4.2 Fr | • IVUS: 45 MHz • NIRF: 750 nm | • IVUS: 4 mm • NIRF: 2 mm | IVUS: N/A | 30/s | Large probe size |
Abran et al. (64) (Coronary) | 4.5 Fr | • IVUS: 40 MHz • NIRF: 750 nm | • IVUS: 4 mm • NIRF: 2 mm | IVUS: 150 μm | 2.7/s | Large probe size | |
Abran et al. (64) (Peripheral) | 9.0 Fr | • IVUS: 15 MHz• NIRF: 750 nm | • IVUS: 4 mm • NIRF: 2 mm | IVUS: 270 μm | 2.7/s | Large probe size | |
NIRF-IVUS-OCT | Maehara et al. (65) | 3.9 Fr | • OCT: 1,310 nm • IVUS: 40 MHz • NIRF: 785 nm | • OCT: 3–5 mm • IVUS: 5–6 mm NIRF: 3–5 mm | – | 20/s | Large probe size |
IVUS, intravenous ultrasound; NIRF, near-infrared fluorescence; OCT, optical coherence tomography.