. 2020 Jun 25;49(8):20200039. doi: 10.1259/dmfr.20200039

Table 5.

Estimated means and standard deviations of volumetric distortion according to MAR algorithm and windowing adjustment obtained for the two CBCT units

CBCT unit Material MAR Windowing
W1 (SD) W2 (SD) p valuea
Picasso Am Off 12.45 (4.73) 26.04 (3.41) p < 0.01
On 12.28 (2.48) 20.52 (2.47) p < 0.01
p valueb 1.00 0.14
CoCr Off 10.37 (8.38) 15.79 (4.06) 0.16
On 6.96 (2.79) 11.58 (2.4) 0.43
p valueb 0.9 0.62
Gu Off 1.15 (4.87) 19.28 (6.76) p < 0.01
On −12.88 (6.33) −7.73 (5.03) 0.23
p valueb p < 0.05 p < 0.05
Ti Off 3.84 (7.85) 19.35 (4.97) p < 0.01
On −11.58 (2) −7.1 (3.24) 0.5
p valueb p < 0.01 p < 0.01
Zi Off 10.9 (7.46) 23.44 (4.12) p < 0.01
On 6.12 (1.76) 12.81 (1.29) p < 0.05
p valueb 0.37 1.00
OP300 Am Off 41.01 (2.42) 47.93 (4.7) 0.14
On 53.37 (5.65) 66.99 (3.85) p < 0.01
p valueb p < 0.01 p < 0.01
CoCr Off 35.22 (5.29) 39.74 (5.29) 0.86
On 52.84 (5.03) 62.09 (2.97) p < 0.01
p valueb p < 0.01 p < 0.01
Gu Off 27.81 (5.34) 37.86 (4.35) p < 0.01
On 28.68 (5.35) 39.91 (7.69) p < 0.01
p valueb 1.00 1.00
Ti Off 28.31 (4.57) 40.8 (2.49) p < 0.01
On 33.83 (12.67) 48.79 (7.47) p < 0.01
p valueb 0.54 p < 0.01
Zi Off 36.78 (2.02) 47.81 (3.51) p < 0.01
On 50.34 (5.28) 63.56 (5.35) p < 0.01
p valueb p < 0.01 p < 0.01

Am, amalgam alloy; CBCT, cone-beam computed tomography; CoCr, cobalt-chromium; Gu, gutta-percha; MAR, metal artefact reduction; SD, standard deviation; Ti, titanium; W1, large window width and upper window level; W2, narrow window width and low window level; Zi, zirconium.

Bold p-values indicate statistically significant differences (p < 0.05).

a

p-value corresponding to comparisons between W1 and W2 considering the same MAR condition.

b

pvalue corresponding to comparisons between MAR algorithm conditions considering the same windowing adjustment.