Abstract
Objectives:
The aim of this study was to evaluate and summarise features of currently and formerly available cone beam CT (CBCT) devices from 1996 to 2019. Additionally, a recommendation for standardised reporting of CBCT characteristics was provided.
Methods and materials:
Information about the features of all available CBCT devices was obtained from the manufacturers’ available data. Moreover, site visits to newly developed CBCT machines’ manufacturers were performed in order to obtain relevant information.
Results:
A total of 279 CBCT models from 47 manufacturers located in 12 countries (Brazil, China, Denmark, Finland, France, Germany, Italy, Japan, Republic of Korea, Slovakia, Thailand, and USA) could be listed. Overall, wide variations in CBCT features and technical specifications were identified.
Conclusions:
CBCT in dentomaxillofacial radiology is a generic term applicable to a broad range of CBCT machines and features. Experimental outcomes and literature statements regarding radiation doses, imaging performance and diagnostic applicability of dental CBCT cannot be simply transferred from one CBCT model to another considering a wide variation in technical characteristics and clinical diagnostic performance. The information tabulated in the present study will be later provided on the International Association of DentoMaxilloFacial Radiology website (www.iadmfr.one).
Keywords: radiology, three-dimensional imaging, cone-beam computed tomography, tomography scanners
Introduction
Dentistry has witnessed tremendous advances over the past decade, with a shift towards digital dentistry, not only for diagnostics but also for surgical planning and treatment. In dentomaxillofacial diagnostics, traditional radiologic examinations are usually limited to two-dimensional (2D) views such as intraoral and panoramic radiographies.1,2 For routine diagnostics, these imaging modalities may often suffice. However, the evolution of diagnostics and treatment in different dental disciplines has raised the need for three-dimensional (3D) imaging to overcome anatomical overlap and distortions inherent to 2D radiology.1,3,4
Technological advances in radiological imaging have led to the introduction of new 3D imaging methods in many fields of radiology, including dentomaxillofacial radiology.5 CT was the first technology to allow visualisation of both hard and soft tissues of the facial bones through the ability to acquire multiple consecutive cross-sectional images.5 Since its introduction, CT imaging of the orofacial region has been of interest as a clinical tool. Since the late 1980s, CT was used for surgical planning in oral and maxillofacial surgery and early endosseous implant placement.6–8 Considering high radiation dose, costs, and space requirements for CT, dedicated devices for 3D imaging in oral healthcare were developed in the 1990s: dental cone beam CT (CBCT),5,9,10 which was available on the market in the late 1990s.1,4,11 This modality may sometimes be denoted as digital volume tomography (DVT), particularly in German-speaking regions.12 Prior to dental CBCT, this imaging modality was applied in medical radiology for angiography, and it is currently used for different other applications such as otorhinolaryngologic (ears, nose and throat—ENT), interventional and intraoperative imaging, cardiac imaging, radiotherapy, musculoskeletal (incl. extremities and spine), breast and peripheral bone imaging.1,3,12–16
Since its introduction in dentomaxillofacial radiology, CBCT has become a widely accepted radiographic tool for diagnosis and treatment planning in oral healthcare.2 During the past two decades, the number of CBCT models has increased markedly, with new models being developed and released continuously.1,17 In 2008, 23 different CBCT devices were reported,18 which increased to 43 devices in 2013.1 Meanwhile, the number of devices increased dramatically along with its increased application in all dental specialties. CBCT is now widespread used and surely not only in relation to diagnosis, yet also in relation to presurgical planning, dental treatment and post-surgical follow-up (such as with grafting procedures).
Dental CBCT devices exhibit wide variability in terms of physical dimensions, features, imaging modalities available, and essential parameters such as X-ray source specifications and exposure parameters, field of view (FOV), and imaging geometry.19 At this moment, there is no recent overview available of all available systems. Therefore, the aim of the present study is to evaluate and summarise currently available CBCT devices and their features. Additionally, a recommendation for standardised reporting of CBCT characteristics was provided
Methods and materials
Medical and grey literature about CBCT devices were reviewed. First, previously published reviews were analysed to assemble CBCT devices used for dentomaxillofacial applications reported in medical literature.1,5,18,20–22 Afterwards, information of CBCT devices that were not found in medical literature was gathered using a search engine. All CBCT devices on the market between 1996 and 2019 were included in this study.
Information about the features of each device was obtained from published articles, available data from manufacturers’ websites, manuals, and brochures. Most of the missing or incomplete data were completed and verified at the International Dental Show in Cologne, Germany, in 2017 and 2019 and at the European Congress of Radiology in Vienna, Austria, in 2018 and 2019, as well as at the European Congress of DentoMaxilloFacial Radiology in Luzern, Switzerland, in 2018.
Features collected about all these devices were:
Manufacturer.
Country of manufacturing.
CBCT Series.
CBCT Model (machines with the optional cephalometric attachment were considered as different models, as well those with different optional fields of view).
Availability on the market.
Imaging modalities available: CBCT only, 2-in-1 (CBCT and panoramic radiography), 3-in-1 (CBCT, panoramic and cephalometric radiography).
Dimensions of width, depth, and height, in meters.
Weight in kilograms.
Patient position condition: standing, sitting (i.e. built-in chair) or supine.
Software available with the CBCT unit.
Kilovoltage range (kV).
Milliamperage range (mA).
Focal spot size in millimetres.
Detector type.
Greyscale/contrast resolution in bits.
FOV size, divided in three categories (small, medium and large) according to Bornstein et al.23 To define the categories, FOV height was multiplied by FOV diameter, representing the FOV surface in cm2. A small FOV was considered when the calculated surface was ≤ 40 cm2; a medium FOV was considered if the surface was > 40 cm2 and ≤ 100 cm2; and a large FOV if it was > 100 cm2.
Use of stitching to acquire extended FOV.
Voxel size in millimetres.
Scan time (i.e. time between first and last projection) in seconds.
Reconstruction time in seconds.
Beam: pulsed or not pulsed (continuous).
Results
From the 143 CBCT series gathered in the present overview, a total of 279 CBCT models were examined, 203 of which available on the market at the time of writing. They are/were produced by 47 manufacturers from 12 countries (Brazil, China, Denmark, Finland, France, Germany, Italy, Japan, Republic of Korea, Slovakia, Thailand, and USA). Republic of Korea had the highest number of CBCT models (n = 72), followed by Japan, Finland, Italy, USA, and France. Brazil, Denmark, and Thailand had a single CBCT series, Eagle 3D, X1, and DentiiScan 2.0 respectively (Figure 1). Vatech (Republic of Korea), which produces 28 models (23 still available on the market) and Planmeca (Finland), which produces 18 models, showed the most variety of CBCT types by individual companies. KaVo, being the owner of different imaging companies (i.e. Gendex, Imaging Science International, Instrumentarium and Soredex) has 32 different CBCT models under its umbrella.
Most of the CBCT models were 2-in-1 models (n = 111), followed by 3-in-1 models (n = 106) while only 62 were CBCT-only models (Figure 2). Two models (RayScan m + SC and DR, Ray Medical Co., Republic of Korea) are 2-in-1 CBCT devices but instead of CBCT and panoramic radiography, they are equipped with CBCT and cephalometric radiography, as they are dedicated to otorhinolaryngology. The machines dimensions vary from 0.8 to 2.5 m in width, 0.8 to 3 m in depth and 1.5 to 2.5 m in height. The main difference in machines size are for 3-in-1 models, which present mean width of 1.9 m. The weight of CBCT devices ranged between 66 and 950 kg and referred only to the devices themselves, without separate workstations (Table 1). In 80% of the CBCT devices, patients are standing while the image is acquired (mostly with wheelchair accessibility), and only 3% of the devices had the patients in supine position (Figure 3).
Table 1.
Manufacturer | Model | Availability on the market | Imaging modalities | Maximum dimensions (width x depth x height) in meters | Maximum weight in kilograms | Patient position | Software | ||
---|---|---|---|---|---|---|---|---|---|
CBCT | PANO | CEPH | |||||||
3M IMTEC (USA) | ILUMA SFOV | No | x | 2 × 1.1 x 2.2 | 350 | Sitting | IlumaVision | ||
ILUMA LFOV | No | x | 2 × 1.1 x 2.2 | 350 | Sitting | IlumaVision | |||
3Shape (Denmark) | 3Shape X1 | Yes | x | x | x | 1.3 × 1.6 x 2.3 | 230 | Standing | 3D Shape viewer |
Acteon (France) | X-Mind Trium | Yes | x | x | 1.1 × 1.3 x 2.4 | 185 | Standing | Acteon Imaging Suite 3D | |
X-Mind Trium (Ceph) | Yes | x | x | x | 1.7 × 1.3 x 2.4 | 215 | Standing | Acteon Imaging Suite 3D | |
X-Mind Trium “True low dose” | Yes | x | x | x | 1.7 × 1.3 x 2.4 | 215 | Standing | Acteon Imaging Suite 3D | |
X-Mind Prime 3D | Yes | x | x | 1 × 1.1 x 2.2 | 67 | Standing | Acteon Imaging Suite 3D | ||
WhiteFox | Yes | x | 1.6 × 1.9 x 2.5 | 275 | Standing | WhiteFox Imaging | |||
Air Techniques (USA) | ProVecta 3D Prime | Yes | x | x | 1.2 × 1.5 x 2.2 | 180 | Standing | VisionX | |
Asahi Roentgen (Japan) | Alphard 2520 | No | x | 2 × 1.7 x 2 | 480 | Sitting | Neo 3D | ||
Alphard 3030 VEGA | No | x | 2 × 1.7 x 2 | 480 | Sitting | Neo 3D | |||
Alioth | No | x | x | 1.2 × 1.4 x 2.3 | 280 | Standing | ADR Plus | ||
Alioth CM | No | x | x | x | 1.9 × 1.4 x 2.3 | 314 | Standing | ADR Plus | |
Auge Zio | No | x | x | 1 × 1.3 x 2.3 | 298 | Standing | ADR Plus | ||
Auge Zio CM | No | x | x | x | 1.9 × 1.3 x 2.3 | 338 | Standing | ADR Plus | |
Auge Zio CM Maxim | No | x | x | x | 1.9 × 1.3 x 2.3 | 348 | Standing | ADR Plus | |
Auge X Zio | No | x | x | 1 × 1.3 x 2.3 | 298 | Standing | ADR Plus | ||
Auge X Zio CM | No | x | x | x | 1.9 × 1.3 x 2.3 | 338 | Standing | ADR Plus | |
Auge X Zio CM Maxim | No | x | x | x | 1.9 × 1.3 x 2.3 | 348 | Standing | ADR Plus | |
Auge Solio Z | Yes | x | x | 1.2 × 1.4 x 2.3 | 200 | Standing | NEOSMART | ||
Auge Solio Z CM | Yes | x | x | x | 1.9 × 1.4 x 2.3 | 220 | Standing | NEOSMART | |
Solio X | Yes | x | x | 1 × 1.3 x 2.3 | 177 | Standing | ? | ||
Solio X Z Maxim | Yes | x | x | x | 1.9 × 1.2 x 2.3 | 197 | Standing | ? | |
PSR 9000N | No | x | ? | ? | Sitting | ? | |||
Biolase (USA) | DaVinci Imaging D3D | No | x | 1.5 × 2.4 x 1.7 | 360 | Supine | ? | ||
Carestream (France) | CS 8100 3D | Yes | x | x | 1.2 × 1.4 x 2.4 | 127 | Standing | Carestream Dental Imaging Software | |
CS 8100SC 3D | Yes | x | x | x | 2 × 1.4 x 2.4 | 127 | Standing | Carestream Dental Imaging Software | |
CS 9000 3D / Kodak 9000 3D | No | x | x | 1.2 × 1.6 x 2.4 | 160 | Standing | ? | ||
CS 9000C 3D / Kodak 9000C 3D | No | x | x | x | 2.2 × 1.6 x 2.4 | 199 | Standing | ? | |
CS 9300 | Yes | x | x | 1.2 × 1.6 x 2.4 | 160 | Standing | Carestream Dental Imaging Software | ||
CS 9300C | Yes | x | x | x | 2.1 × 1.6 x 2.4 | 199 | Standing | Carestream Dental Imaging Software | |
CS 9300 Select | Yes | x | x | 1.2 × 1.6 x 2.4 | 199 | Standing | Carestream Dental Imaging Software | ||
CS 9300C Select | Yes | x | x | x | 2.1 × 1.6 x 2.4 | 199 | Standing | Carestream Dental Imaging Software | |
CS 9300 ENT | Yes | x | 1.2 × 1.6 x 2.4 | 160 | Standing | Carestream Dental Imaging Software | |||
Kodak 9500 MFOV | No | x | 1.7 × 1.7 x 2.3 | 176 | Standing | Kodak Dental Imaging | |||
Kodak 9500 LFOV | No | x | 1.7 × 1.7 x 2.3 | 176 | Standing | Kodak Dental Imaging | |||
CS 9600 12 × 10 | Yes | x | x | 1.3 × 1.7 x 2.5 | 210 | Standing | CS Imaging | ||
CS 9600 16 × 10 | Yes | x | x | 1.3 × 1.7 x 2.5 | 210 | Standing | CS Imaging | ||
CS 9600 16 × 17 | Yes | x | x | 1.3 × 1.7 x 2.5 | 210 | Standing | CS Imaging | ||
Castellini (Italy) | X Radius Compact 3D | No | x | x | 0.9 × 1.1 x 2.3 | 90 | Standing | iRYS | |
X Radius Trio SFOV | No | x | x | 1.3 × 1.5 x 2.5 | 170 | Standing | iRYS | ||
X Radius Trio SFOV (Ceph) | No | x | x | x | 1.8 × 1.5 x 2.5 | 190 | Standing | iRYS | |
X Radius Trio LFOV | No | x | x | 1.3 × 1.5 x 2.5 | 170 | Standing | iRYS | ||
X Radius Trio LFOV (Ceph) | No | x | x | x | 1.8 × 1.5 x 2.5 | 190 | Standing | iRYS | |
Dürr (Germany) | VistaVOX S | Yes | x | x | 1 × 1.3 x 2.5 | 180 | Standing | VistaSoft | |
VistaVOX S (Ceph) | Yes | x | x | x | 1.5 × 1.2 x 2.2 | 202 | Standing | VistaSoft | |
Dabi Atlante (Brazil) | Eagle 3D | Yes | x | x | 1.9 × 1.8 x 2.5 | 115 | Standing | OnDemand 3D Dental | |
Eagle 3D (Ceph) | Yes | x | x | x | 2.5 × 1.8 x 2.5 | 152 | Standing | OnDemand 3D Dental | |
Dentium (Republic of Korea) | Rainbow CT | Yes | x | 1.9 × 1.5 x 2.3 | 150 | Standing | DaVinci | ||
Denstply Sirona (Germany) | Orthophos S 3D | No | x | x | 1.3 × 1.4 x 2.3 | 110 | Standing | Sidexis 4 | |
Orthophos S 3D (Ceph) | No | x | x | x | 2.2 × 1.4 x 2.3 | 132 | Standing | Sidexis 4 | |
Orthophos SL 3D | Yes | x | x | 1.3 × 1.4 x 2.3 | 110 | Standing | Sidexis 4 | ||
Orthophos SL 3D (Ceph) | Yes | x | x | x | 2.2 × 1.4 x 2.3 | 132 | Standing | Sidexis 4 | |
Orthophos XG 3D | Yes | x | x | 1.3 × 1.4 x 2.3 | 110 | Standing | Sidexis 4 | ||
Orthophos XG 3D (Ceph) | Yes | x | x | x | 2.2 × 1.4 x 2.3 | 132 | Standing | Sidexis 4 | |
GALILEOS Compact | No | x | 1.6 × 1.6 x 2.3 | 140 | Standing | Sidexis 4 | |||
GALILEOS Comfort | No | x | 1.6 × 1.6 x 2.3 | 140 | Standing | Sidexis 4 | |||
GALILEOS Comfort Plus | Yes | x | 1.6 × 1.6 x 2.3 | 140 | Standing | Sidexis 4 | |||
Fussen (China) | Dentrix 20 | Yes | x | x | x | 1.7 × 1.1 x 2.3 | 160 | Standing | Fussen DenView |
FONA Dental (Slovakia) | Stellaris 3D | Yes | x | x | 1.4 × 1.1 x 2.3 | 109 | Standing | Stellaris PRO | |
Stellaris 3D (Ceph) | Yes | x | x | x | 2.4 × 1.1 x 2.3 | 137 | Standing | Stellaris PRO | |
FONA X PAN 3D | Yes | x | x | ? | 98 | Standing | FONA OrisWin | ||
FONA X PAN 3D Plus | Yes | x | x | x | ? | 125 | Standing | FONA OrisWin | |
Genoray (Republic of Korea) | Volux 6 | No | x | 1.2 × 1.3 x 2 | 220 | Sitting | ? | ||
Volux 9 (Dental/ENT) | No | x | 1.5 × 1.4 x 1.9 | 220 | Sitting | ? | |||
Volux 21 | No | x | x | 0.9 × 1.5 x 2.4 | 250 | Standing | TRIANA | ||
Volux 21C | No | x | x | x | 2.1 × 1.5 x 2.3 | 300 | Standing | TRIANA | |
Papaya 3D | Yes | x | x | 1.2 × 1.1 x 2.4 | ? | Standing | TRIANA | ||
Papaya 3D Plus | Yes | x | x | x | 1.9 × 1.1 x 2.4 | ? | Standing | TRIANA | |
Papaya 3D Premium NV | Yes | x | x | 1.3 × 1.4 x 2.4 | ? | Standing | TRIANA | ||
Papaya 3D Premium NV (Ceph) | Yes | x | x | x | 2 × 1.4 x 2.4 | ? | Standing | TRIANA | |
Papaya 3D Premium LV | Yes | x | x | 1.3 × 1.4 x 2.4 | ? | Standing | TRIANA | ||
Papaya 3D Premium LV (Ceph) | Yes | x | x | x | 2 × 1.4 x 2.4 | ? | Standing | TRIANA | |
Papaya 3D Premium ENT | Yes | x | 1.2 × 1.1 x 2.4 | ? | Standing | TRIANA | |||
HDX Will (Republic of Korea) | Dentri α Classic | Yes | x | x | 1.2 × 1.4 x 2.5 | 233 | Standing | OnDemand 3D | |
Dentri α Extended | Yes | x | x | 1.2 × 1.4 x 2.5 | 243 | Standing | OnDemand 3D | ||
Dentri Cα Classic | Yes | x | x | x | 2 × 1.4 x 2.5 | 243 | Standing | OnDemand 3D | |
Dentri Cα Extended | Yes | x | x | x | 2 × 1.4 x 2.5 | 260 | Standing | OnDemand 3D | |
Dentri Sα Classic | Yes | x | x | x | 1.9 × 1.4 x 2.5 | 260 | Standing | OnDemand 3D | |
Dentri Sα Extended | Yes | x | x | x | 1.9 × 1.4 x 2.5 | 270 | Standing | OnDemand 3D | |
Dinnova 3 2520D/3030D | Yes | x | x | x | 1.5 × 1.7 x 1.9 | 480 | Sitting | OnDemand 3D | |
Q-Face (non-stitch) | Yes | x | x | 1.7 × 1.7 x 2.4 | 285 | Standing | ? | ||
Q-Face (one stitch) | Yes | x | x | 1.7 × 1.7 x 2.4 | 305 | Standing | ? | ||
Q-Face (two stitch) | Yes | x | x | 1.7 × 1.7 x 2.5 | 305 | Standing | ? | ||
Q-Face-S (non-stitch) | Yes | x | x | x | 1.9 × 1.7 x 2.4 | 310 | Standing | ? | |
Q-Face-S (one stitch) | Yes | x | x | x | 1.9 × 1.7 x 2.4 | 330 | Standing | ? | |
Q-Face-S (two stitch) | Yes | x | x | x | 1.9 × 1.7 x 2.5 | 330 | Standing | ? | |
Hitachi (Japan) | CB Mercuray | No | x | 1.8 × 1.9 x 2.3 | 950 | Sitting | CB Works | ||
CB Throne | No | x | 1.8 × 1.8 x ? | ? | Sitting | CB Works | |||
ImageWork (USA) | Panoura 18S | Yes | x | x | 1 × 1.1 x 2.3 | 165 | Standing | OnDemand 3D | |
Panoura 18S (Ceph) | Yes | x | x | x | 1.9 × 1.1 x 2.3 | 205 | Standing | OnDemand 3D | |
J Morita (Japan) | 3D Accuitomo | No | x | 1.6 × 1.2 x 2.1 | 400 | Sitting | i-Dixel | ||
3D Accuitomo FPD | No | x | 1.6 × 1.2 x 2.1 | 400 | Sitting | i-Dixel | |||
3D Accuitomo FPD 80 | No | x | 1.6 × 1.2 x 2.1 | 400 | Sitting | i-Dixel | |||
3D Accuitomo 170 | Yes | x | 1.6 × 1.2 x 2.1 | 400 | Sitting | i-Dixel | |||
Veraview X800 F40P | Yes | x | x | 1.4 × 1.2 x 2.3 | 185 | Standing | i-Dixel | ||
Veraview X800 F40CP | Yes | x | x | x | 2.0 × 1.2 x 2.3 | 220 | Standing | i-Dixel | |
Veraview X800 R100P | Yes | x | x | 1.4 × 1.2 x 2.3 | 185 | Standing | i-Dixel | ||
Veraview X800 R100CP | Yes | x | x | x | 2.0 × 1.2 x 2.3 | 220 | Standing | i-Dixel | |
Veraview X800 F150P | Yes | x | x | 1.4 × 1.2 x 2.3 | 185 | Standing | i-Dixel | ||
Veraview X800 F150CP | Yes | x | x | x | 2.0 × 1.2 x 2.3 | 220 | Standing | i-Dixel | |
Veraviewepocs 3D F40 | No | x | x | 1 × 1.3 x 2.4 | 190 | Standing | i-Dixel | ||
Veraviewepocs 3D F40CP | No | x | x | x | 2 × 1.3 x 2.4 | 260 | Standing | i-Dixel | |
Veraviewepocs 3D R100 | Yes | x | x | 1 × 1.3 x 2.4 | 190 | Standing | i-Dixel | ||
Veraviewpocs 3D R100CP | Yes | x | x | x | 2 × 1.3 x 2.4 | 260 | Standing | i-Dixel | |
KaVo Kerr (USA/Finland) | KaVo 3D eXam | No | x | 1.2 × 0.8 x 1.8 | ? | Sitting | DTX Studio | ||
KaVo OP 3D | Yes | x | x | 0.8 × 1.1 x 2.5 | 120 | Standing | DTX Studio | ||
KaVo OP 3D (Ceph) | Yes | x | x | x | 1.5 × 1.7 x 2.5 | 155 | Standing | DTX Studio | |
KaVo OP 3D Pro Small Panel | Yes | x | x | 1 × 1.4 x 2.4 | 200 | Standing | DTX Studio | ||
KaVo OP 3D Pro Small Panel (Ceph) | Yes | x | x | x | 2 × 1.4 x 2.4 | 250 | Standing | DTX Studio | |
KaVo OP 3D Pro Large Panel | Yes | x | x | 1 × 1.4 x 2.4 | 200 | Standing | DTX Studio | ||
KaVo OP 3D Pro Large Panel (Ceph) | Yes | x | x | x | 2 × 1.4 x 2.4 | 250 | Standing | DTX Studio | |
KaVo OP 3D Vision (V8) | Yes | x | 1.2 × 1.3 x 1.8 | 230 | Sitting | DTX Studio | |||
KaVo OP 3D Vision (V10) | Yes | x | 1.2 × 1.3 x 1.8 | 230 | Sitting | DTX Studio | |||
KaVo OP 3D Vision (V17) | Yes | x | 1.2 × 1.3 x 1.8 | 230 | Sitting | DTX Studio | |||
KaVo (Gendex - USA) | GXDP-800 | No | x | x | 1.1 × 1.4 x 2.4 | 200 | Standing | ? | |
GXDP-800 (Ceph) | No | x | x | x | 2 × 1.4 x 2.4 | 250 | Standing | ? | |
GXDP-700-S | Yes | x | x | 1.1 × 1.4 x 2.5 | 200 | Standing | InVivo 5 | ||
GXDP-700-SC | Yes | x | x | x | 2 × 1.4 x 2.5 | 250 | Standing | InVivo 5 | |
GXCB-500 | Yes | x | 1.2 × 1.3 x 1.8 | 230 | Standing | InVivo5 | |||
KaVo (Imaging Science International - USA) | i-CAT Classic | No | x | 1.1 × 1.3 x 1.8 | 192 | Sitting | i-CAT Vision | ||
i-CAT Precise | No | x | 1.2 × 1.3 x 1.8 | 231 | Sitting | i-CAT Vision | |||
i-CAT Next Generation | No | x | 1.2 × 0.9 x 1.8 | 231 | Sitting | i-CAT Vision | |||
i-CAT FLX MV | No | x | 1.2 × 0.9 x 1.8 | 230 | Sitting | i-CAT Vision | |||
i-CAT FLX V8 | Yes | x | 1.2 × 1.3 x 1.8 | 230 | Sitting | DTX Studio | |||
i-CAT FLX V10 | Yes | x | 1.2 × 1.3 x 1.8 | 230 | Sitting | DTX Studio | |||
i-CAT FLX V17 | Yes | x | 1.2 × 1.3 x 1.8 | 230 | Sitting | DTX Studio | |||
KaVo (Instrumentarium – Finland) | OP200 D | No | x | x | 2 × 1.3 x 2.3 | 175 | Standing | ? | |
OC 200 D | No | x | x | x | 2 × 1.3 x 2.3 | 210 | Standing | ? | |
OP300 | Yes | x | x | x | 2 × 1.4 x 2.5 | 250 | Standing | OnDemand 3D | |
OP300 Maxio | Yes | x | x | x | 2 × 1.4 x 2.5 | 250 | Standing | OnDemand 3D | |
KaVo (Soredex – Finland) | SCANORA 3D | No | x | x | 1.6 × 1.4 x 2 | 310 | Sitting | OnDemand 3D | |
SCANORA 3DX | No | x | x | 1.6 × 1.4 x 2 | 310 | Sitting | OnDemand 3D | ||
CRANEX 3D | No | x | x | 1 × 1.4 x 2.4 | 200 | Standing | OnDemand 3D | ||
CRANEX 3D (Ceph) | No | x | x | x | 2 × 1.4 x 2.4 | 250 | Standing | OnDemand 3D | |
CRANEX 3DX | No | x | x | 1 × 1.4 x 2.4 | 200 | Standing | OnDemand 3D | ||
CRANEX 3DX (Ceph) | No | x | x | x | 2 × 1.4 x 2.4 | 250 | Standing | OnDemand 3D | |
LargeV (China) | HiRes 3D Dental | Yes | x | 1.1 × 1.6 x 2.1 | 340 | Sitting | SmartV | ||
HiRes 3D MAX | Yes | x | 1.1 × 1.6 x 2.1 | 340 | Sitting | SmartV | |||
Smart Dental | Yes | x | x | 1.1 × 1.5 x 2.4 | 285 | Standing | SmartV | ||
Smart Dental (Ceph) | Yes | x | x | x | 2 × 1.5 x 2.4 | 335 | Standing | SmartV | |
MyRay (Italy) | SkyView | No | x | 1.5 × 2.5 x 1.8 | 360 | Supine | SkyView | ||
Hyperion X5 3D/2D | Yes | x | x | 0.9 × 1.1 x 2.3 | 90 | Standing | iRYS | ||
Hyperion X5 3D/2D CEPH | Yes | x | x | x | 1.8 × 1.1 x 2.3 | 115 | Standing | iRYS | |
Hyperion X9 Full FOV | Yes | x | x | 1.3 × 1.5 x 2.5 | 170 | Standing | iRYS | ||
Hyperion X9 Full FOV CEPH | Yes | x | x | x | 1.8 × 1.5 x 2.5 | 190 | Standing | iRYS | |
Hyperion X9 Extended FOV | Yes | x | x | 1.3 × 1.5 x 2.5 | 170 | Standing | iRYS | ||
Hyperion X9 Extended FOV CEPH | Yes | x | x | x | 1.8 × 1.5 x 2.5 | 190 | Standing | iRYS | |
Hyperion X9 Pro 10 × 8 version | Yes | x | x | 1.3 × 1.5 x 2.5 | 155 | Standing | iRYS | ||
Hyperion X9 Pro CEPH 10 × 8 version | Yes | x | x | x | 1.8 × 1.5 x 2.5 | 175 | Standing | iRYS | |
Hyperion X9 Pro 13 × 16 version | Yes | x | x | 1.3 × 1.5 x 2.5 | 155 | Standing | iRYS | ||
Hyperion X9 Pro CEPH 13 × 16 version | Yes | x | x | x | 1.8 × 1.5 x 2.5 | 175 | Standing | iRYS | |
Meyer (China) | SS-91010D Pro-3D | Yes | x | x | 1.1 × 1.4 x 2.5 | ? | Standing | DCTViewer | |
SS-91010D Pro-3De | Yes | x | x | x | 1.9 × 1.4 x 2.5 | ? | Standing | DCTViewer | |
X12008D Pro-3D | Yes | x | 1.5 × 1.7 x 2 | 350 | Sitting | DCTViewer | |||
PiXAMED (Thailand) | DentiiScan 2.0 | Yes | x | 1 × 1.3 x 2.4 | 190 | Standing | In house software | ||
Owandy (France) | I-MAX 3D | Yes | x | x | 1 × 1.1 x 2.2 | 66 | Standing | QuickVision | |
I-MAX 3D Touch | Yes | x | x | 1.2 × 1.3 x 2.5 | 161 | Standing | QuickVision | ||
I-MAX 3D Touch CEPH | Yes | x | x | x | 2 × 1.3 x 2.5 | 186 | Standing | QuickVision | |
Osstem (Republic of Korea) | Osstem Implant CBCT T1 | Yes | x | x | x | 1.9 × 1.2 x 2.3 | 210 | Standing | Oneclinic |
Planmeca (Finland) | ProMax 3D Classic | Yes | x | x | 1.2 × 1.3 x 2.4 | 113 | Standing | Romexis | |
ProMax 3D Classic (Ceph) | Yes | x | x | x | 2 × 1.3 x 2.4 | 128 | Standing | Romexis | |
ProMax 3D S | Yes | x | x | 1.2 × 1.3 x 2.4 | 113 | Standing | Romexis | ||
ProMax 3D S (Ceph) | Yes | x | x | x | 2 × 1.3 x 2.4 | 128 | Standing | Romexis | |
ProMax 3D Plus | Yes | x | x | 1.2 × 1.4 x 2.4 | 131 | Standing | Romexis | ||
ProMax 3D Plus (Ceph) | Yes | x | x | x | 2.1 × 1.4 x 2.4 | 146 | Standing | Romexis | |
ProMax 3D Plus ENT | Yes | x | x | 1.2 × 1.4 x 2.4 | 131 | Standing | Romexis | ||
ProMax 3D Plus ENT (Ceph) | Yes | x | x | x | 2.1 × 1.4 x 2.4 | 146 | Standing | Romexis | |
ProMax 3D Mid | Yes | x | x | 1.2 × 1.4 x 2.4 | 131 | Standing | Romexis | ||
ProMax 3D Mid (Ceph) | Yes | x | x | x | 2.1 × 1.4 x 2.4 | 146 | Standing | Romexis | |
ProMax 3D Mid ENT | Yes | x | x | 1.2 × 1.4 x 2.4 | 131 | Standing | Romexis | ||
ProMax 3D Mid ENT (Ceph) | Yes | x | x | x | 2.1 × 1.4 x 2.4 | 146 | Standing | Romexis | |
ProMax 3D Max | Yes | x | x | 1.2 × 1.4 x 2.4 | 131 | Standing | Romexis | ||
ProMax 3D Max ENT | Yes | x | x | 1.2 × 1.4 x 2.4 | 131 | Standing | Romexis | ||
Viso | Yes | x | x | 1.3 × 1.5 x 2.3 | 165 | Standing | Romexis | ||
Viso (Ceph) | Yes | x | x | x | 2.1 × 1.5 x 2.4 | 180 | Standing | Romexis | |
Viso (Vertical stitching) | Yes | x | x | 1.3 × 1.5 x 2.3 | 165 | Standing | Romexis | ||
Viso (Vertical stitching) (Ceph) | Yes | x | x | x | 2.1 × 1.5 x 2.4 | 180 | Standing | Romexis | |
Pointnix (Republic of Korea) | Point I3D | Yes | x | 1.2 × 1.1 x 1.8 | Sitting | RealScan | |||
Point 3D Combi 500 | Yes | x | x | 1.1 × 1.3 x 2.3 | 150 | Standing | Romexis | ||
Point 3D Combi 500c | Yes | x | x | x | 2.1 × 1.3 x 2.3 | 190 | Standing | Romexis | |
Point 3D Combi 500 s | Yes | x | x | x | 2.1 × 1.3 x 2.3 | 185 | Standing | Romexis | |
Point 800 HD Plus | Yes | x | x | 1.1 × 1.3 x 2.3 | 150 | Standing | Romexis | ||
Point 800c HD Plus | Yes | x | x | x | 2.1 × 1.3 x 2.3 | 190 | Standing | Romexis | |
Point 800 s HD Plus | Yes | x | x | x | 2.1 × 1.3 x 2.3 | 185 | Standing | Romexis | |
PreXion (Japan) | PreXion 3D | No | x | 1.2 × 1.6 x 2 | 400 | Sitting | Prexion3D Viewer | ||
PreXion 3D Elite | Yes | x | x | 1.2 × 1.6 x 2 | 390 | Sitting | Prexion3D Viewer | ||
PreXion 3D Elite Element | Yes | x | x | 1.2 × 1.6 x 2 | 390 | Sitting | Prexion3D Viewer | ||
PreXion 3D Eclipse | Yes | x | x | 1.2 × 1.3 x 2 | 260 | Sitting | Prexion3D Viewer | ||
PreXion 3D Eclipse (Ceph) | Yes | x | x | x | 1.8 × 1.3 x 2 | 300 | Sitting | Prexion3D Viewer | |
PreXion 3D Excelsior Ex | Yes | x | x | 1 × 1.2 x 2.2 | 165 | Standing | Prexion3D Viewer | ||
PreXion 3D Excelsior Ex (Ceph) | Yes | x | x | x | 1.8 × 1.2 x 2.2 | 200 | Standing | Prexion3D Viewer | |
PreXion 3D Excelsior Plus | Yes | x | x | 1 × 1.2 x 2.2 | 165 | Standing | Prexion3D Viewer | ||
PreXion 3D Excelsior Plus (Ceph) | Yes | x | x | x | 1.8 × 1.2 x 2.2 | 200 | Standing | Prexion3D Viewer | |
PreXion 3D Excelsior Pro | Yes | x | x | 1 × 1.2 x 2.2 | 165 | Standing | Prexion3D Viewer | ||
PreXion 3D Excelsior Pro (Ceph) | Yes | x | x | x | 1.8 × 1.2 x 2.2 | 200 | Standing | Prexion3D Viewer | |
PreXion 3D Explorer | Yes | x | x | x | 1.1 × 1.6 x 2.3 | 185 | Standing | Prexion3D Viewer | |
Quantitative Radiology / Cefla Dental Group (Italy) | GiANO HR Prime | Yes | x | x | 1.2 × 1.4 x 2.5 | 155 | Standing | NNT | |
GiANO HR Prime (Ceph) | Yes | x | x | x | 1.8 × 1.2 x 2.5 | 175 | Standing | NNT | |
GiANO HR Advanced | Yes | x | x | 1.2 × 1.4 x 2.5 | 155 | Standing | NNT | ||
GiANO HR Advanced (Ceph) | Yes | x | x | x | 1.8 × 1.2 x 2.5 | 175 | Standing | NNT | |
GiANO HR Professional | Yes | x | x | 1.2 × 1.4 x 2.5 | 155 | Standing | NNT | ||
GiANO HR Professional (Ceph) | Yes | x | x | x | 1.8 × 1.2 x 2.5 | 175 | Standing | NNT | |
GiANO/NewTom VG3 | Yes | x | x | 1.4 × 1.4 x 2.4 | 170 | Standing | NNT | ||
GiANO/NewTom VG3 (Ceph) | Yes | x | x | x | 1.9 × 1.4 x 2.4 | 190 | Standing | NNT | |
Go 2D/3D Imaging | Yes | x | x | 0.9 × 1.1 x 2.3 | 90 | Standing | NNT | ||
NewTom 9000 Maxiscan | No | x | 2.5 × 3 x 2.4 | 320 | Supine | ||||
NewTom VG | No | x | 1.1 × 1.5 x 2.3 | 272 | Standing | NNT | |||
NewTom VG Flex | No | x | 1.2 × 1.5 x 2 | 272 | Sitting | NNT | |||
NewTom VGi | No | x | 1.2 × 1.5 x 2.3 | 272 | Standing | NNT | |||
NewTom VGi Flex | No | x | 1.2 × 1.5 x 2 | 272 | Standing | NNT | |||
NewTom VGi evo | Yes | x | x | x | 1.3 × 1.6 x 2.3 | 337 | Standing | NNT | |
NewTom 3G | No | x | 2 × 2.5 x 2 | 480 | Supine | NewTom 3G Expert | |||
NewTom 5G | No | x | 1.8 × 2.3 x 1.8 | 530 | Supine | NNT | |||
NewTom 5G XL | Yes | x | 1.8 × 3.6 x 1.8 | 660 | Supine | NNT | |||
Ritter Imaging (Germany) | Orion | No | x | ? | ? | Sitting | ? | ||
Ray Medical (Republic of Korea) | RAYSCAN Symphony BC | No | x | 1.2 × 1.6 x 2 | 355 | Sitting | ? | ||
RAYSCAN Symphony V | No | x | 1.4 × 1.7 x 2 | 355 | Sitting | ? | |||
RAYSCAN α−3D | Yes | x | x | 1.5 × 1.7 x 2.3 | 150 | Standing | ? | ||
RAYSCAN α−3D (SC, OCS, OCL) | Yes | x | x | x | 2.1 × 1.7 x 2.3 | 177.5 | Standing | ? | |
RAYSCAN α + 130 | Yes | x | x | 1.5 × 1.5 x 2.3 | 148 | Standing | ? | ||
RAYSCAN α + 130 (SC, OCS, OCL) | Yes | x | x | x | 2 × 1.5 x 2.3 | 164 | Standing | ? | |
RAYSCAN α + 160 | Yes | x | x | 1.5 × 1.5 x 2.3 | 148 | Standing | ? | ||
RAYSCAN α + 160 (SC, OCS, OCL) | Yes | x | x | x | 2 × 1.5 x 2.3 | 164 | Standing | ? | |
RAYSCAN m+ | Yes | x | 1.5 × 1.5 x 2.3 | ? | Standing | ? | |||
RAYSCAN m + SC | Yes | x | x | 1.9 × 1.5 x 2.3 | ? | Standing | ? | ||
RAYSCAN m + DR (with chest X-ray) | Yes | x | x | 1.9 × 1.5 x 2.3 | ? | Standing | ? | ||
Ray Medical (Republic of Korea) / Apteryx (USA) | RAYSCAN α-Edge | Yes | x | x | 1.5 × 1.7 x 2.3 | 150 | Standing | ? | |
RAYSCAN α-SM Edge | Yes | x | x | x | 2.1 × 1.7 x 2.3 | 177.5 | Standing | ? | |
Streamhealth Dental (USA) | Trophypan Smart 3D | Yes | x | x | 1.2 × 1.4 x 2.4 | 92 | Standing | ? | |
Suni Medical Imaging (USA) | Suni3D | No | x | x | 1 × 1.2 x 2.3 | ? | Standing | ? | |
Suni3D (Ceph) | No | x | x | x | 1.9 × 1.2 x 2.3 | ? | Standing | ? | |
Suni Q3D | No | x | x | 1.2 × 1.2 x 2.4 | 145 | Standing | Q3D | ||
Suni Q3D (Ceph) | No | x | x | x | 1.9 × 1.2 x 2.4 | 160 | Standing | Q3D | |
Trident (Italy) | X-View Cone Beam | Yes | x | x | 0.9 × 1.1 x 2.2 | 95 | Standing | Deep View/Xelis | |
X-View Cone Beam (Ceph) | Yes | x | x | x | 1.7 × 1.1 x 2.2 | 125 | Standing | Deep View/Xelis | |
Vatech (Republic of Korea) | PaX-i3D | Yes | x | x | 1.1 × 1.4 x 2.4 | 120 | Standing | Ez3D plus | |
PaX-i3D (SC or OP) | Yes | x | x | x | 1.9 × 1.3 x 2.4 | 150 | Standing | Ez3D plus | |
PaX-i3D Smart / i3D Smart 8 × 8 | Yes | x | x | 1.3 × 1.4 x 2.4 | 137 | Standing | Ez3D plus | ||
PaX-i3D Smart (Ceph)/ i3D Smart 8 × 8 SC | Yes | x | x | x | 1.9 × 1.4 x 2.4 | 162 | Standing | Ez3D plus | |
PaX-i3D Smart 2/Green Smart / Smart Plus / i3D Smart | Yes | x | x | 1.1 × 1.4 x 2.4 | 137 | Standing | Ez3D plus | ||
PaX-i3D Smart 2 (Ceph) / Green Smart SC / Smart Plus RC / i3D Smart RC | Yes | x | x | x | 1.9 × 1.3 x 2.4 | 162 | Standing | Ez3D plus | |
PaX-i3D Green / Green CT | Yes | x | x | 1.2 × 1.4 x 2.4 | 137 | Standing | Ez3D plus | ||
PaX-i3DGreen SC or OP / Green CT SC or OP | Yes | x | x | x | 2 × 1.4 x 2.4 | 162 | Standing | Ez3D plus | |
PaX-i3D 2/Green CT 2/Green 16 | Yes | x | x | 1.1 × 1.5 x 2.4 | 187 | Standing | Ez3D plus | ||
PaX-i3D 2 (Ceph) / Green CT 2 (Ceph) / Green 16 (Ceph) | Yes | x | x | x | 1.9 × 1.5 x 2.4 | 212 | Standing | Ez3D plus | |
Green 18 | Yes | x | x | 1.1 × 1.5 x 2.4 | 187 | Standing | Ez3D plus | ||
Green 18 (Ceph) | Yes | x | x | x | 1.9 × 1.5 x 2.4 | 212 | Standing | Ez3D plus | |
Green 21/i3D Premium (AutoCeph) | Yes | x | x | 1.5 × 1.6 x 2.2 | 321 | Sitting | Ez3D plus | ||
PaX-Flex3D P | Yes | x | x | 1 × 1.5 x 2.3 | 185 | Standing | Ez3D plus | ||
PaX-Flex3D PC | Yes | x | x | x | 2 × 1.5 x 2.3 | 225 | Standing | Ez3D plus | |
PaX-Uni3D P | Yes | x | x | 1 × 1.5 x 2.3 | 185 | Standing | Ez3D plus | ||
PaX-Uni3D PC | Yes | x | x | x | 2 × 1.5 x 2.3 | 225 | Standing | Ez3D plus | |
PaX-Duo3D | Yes | x | x | 1.1 × 1.6 x 2.3 | 220 | Standing | Ez3D plus | ||
PaX-Zenith3D | Yes | x | x | 1.8 × 2 x 1.9 | 493 | Sitting | Ez3D plus | ||
PaX-Reve3D | Yes | x | x | 1.4 × 1.6 x 2.3 | 210 | Standing | Ez3D plus | ||
PaX-Reve3D OS | Yes | x | x | x | 2.1 × 1.6 x 2.3 | 250 | Standing | Ez3D plus | |
Picasso Trio | Yes | x | x | 1 × 1.5 x 2.3 | 185 | Standing | Ez3D plus | ||
Picasso Trio (Ceph) | Yes | x | x | x | 2 × 1.5 x 2.3 | 225 | Standing | Ez3D plus | |
Picasso Duo | No | x | x | 0.9 × 1.5 x 2.4 | ? | Standing | Ez3D | ||
Picasso Master 3D | No | x | 1.5 × 1.9 x 2.3 | ? | Sitting | Ex3D | |||
Picasso Pro 3D | No | x | 1.4 × 1.6 x 1.8 | ? | Sitting | Ez3D | |||
PaX 500 ECT | No | x | x | 1 × 1.3 x 2.3 | ? | Standing | ECT Viewer | ||
PaX 500 ECT (Ceph) | No | x | x | x | 1.9 × 1.3 x 2.3 | ? | Standing | ECT Viewer | |
Villa Sistemi Medicalli (Italy) | Rotograph Evo 3D | Yes | x | x | 1.1 × 1.3 x 2.5 | 191 | Standing | Dental Studio | |
Rotograph Evo 3D (Ceph) | Yes | x | x | x | 2 × 1.3 x 2.5 | 216 | Standing | Dental Studio | |
Rotograph Prime 3D | Yes | x | x | 1 × 1.1 x 2.2 | 67 | Standing | Quick Vision / Villa 3D | ||
UEG Medical (China) | Spectral | Yes | x | 1.2 × 0.8 x 2 | 320 | Sitting | UEG RealView | ||
Xoran Technologies (USA) | Minicat | No | x | 0.9 × 1 x 1.7 | 204 | Sitting | Minicat Viewing Solutions | ||
Minicat IQ (ENT) | Yes | x | 0.9 × 1 x 1.7 | 204 | Sitting | Minicat Viewing Solutions | |||
Minicat 2020 | Yes | x | ? | ? | Sitting | ? | |||
XCat (Portable) | Yes | x | 0.8 × 1.2 x 1.5 | 235 | Supine | ? | |||
Yoshida (Japan) | Finecube | No | x | 1.9 × 1.2 x 1.6 | 390 | Sitting | PreXion 3D Viewer | ||
X-era Smart F+ | Yes | x | x | 0.9 × 1.1 x 2.2 | 170 | Standing | ? | ||
X-era Smart F+ (Ceph) | Yes | x | x | x | 1.9 × 1.1 x 2.2 | 210 | Standing | ? | |
Yoshida (Japan) / ImageWorks (USA) | X-era NF / Panoura X-era NF | Yes | x | x | 1.1 × 1.4 x 2.4 | ? | Standing | OneSystem | |
X-era NF (Ceph) / Panoura X-era NF (Ceph) | Yes | x | x | x | 1.9 × 1.4 x 2.3 | ? | Standing | OneSystem | |
X-era MF / Panoura X-era MF | Yes | x | x | 1.1 × 1.4 x 2.3 | ? | Standing | OneSystem | ||
X-era MF (Ceph) / X-era MF (Ceph) | Yes | x | x | x | 1.9 × 1.4 x 2.3 | ? | Standing | OneSystem |
CBCT: cone-beam computed tomography; PANO: panoramic radiography; CEPH: cephalometric radiography.
Regarding exposure parameters, kV and mA values were observed as constant or adjustable; most of the devices had adjustable kV and mA options (Table 2). Machines that used the lowest kV in the found range (as low as 40 kV) are not available on the market anymore. Few devices have fixed kV, while the overall kV range for devices currently on the market is from 50 to 120. It is crucial to note that these kV range do not always specifically refer to the actual selectable range in ‘CBCT mode’; in many cases, especially for 2-in-1 and 3-in-1 units, the low end of the kV range may only apply to the panoramic, cephalometric and/or service mode. Also, for tube current, a vast diversity of ranges is encountered among the CBCT devices, with the overall range being from 1 to 32 mA, which is the case of NewTom VGi EVO (Quantitative Radiology/Cefla Dental Group, Italy). CBCT devices operate with 0.15, 0.2, 0.3, 0.4 0.5, 0.6, and 0.7 mm focal spot size, with the majority being 0.5 mm. NewTom 9000 Maxiscan (the first commercial CBCT machine for dental application) had a focal spot of 1.2 × 0.8 mm.
Table 2.
Manufacturer | Model | kV | mA | Focal Spot (mm) | Detector Typea | Detector Gray Scale (bits) | FOVb | FOV Stitching | Voxel Size (min-max) (mm) | Scan Time (s)c | Reconstruction Time (s) | Pulsed Beam | ||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Small | Medium | Large | ||||||||||||
3M IMTEC (USA) | ILUMA SFOV | 120 | 1–3.8 | 0.3 | FPD (A-Si) | 16 | x | No | 0.09–0.4 | 7.8–40 | >120 | No | ||
ILUMA LFOV | 120 | 1–3.8 | 0.3 | FPD (A-Si) | 16 | x | No | 0.09–0.4 | 7.8–40 | >120 | No | |||
3Shape (Denmark) | X1 | 60–90 | 4–12 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.1–0.4 | 12–14 | 300 | Yes | |
Acteon (France) | X-Mind Trium | 80–90 | 4–10 | 0.5 | FPD (CMOS) | 12 | x | x | No | 0.075–0.15 | 12–30 | 29 | Yes | |
X-Mind Trium (Ceph) | 80–90 | 4–10 | 0.5 | FPD (CMOS) | 12 | x | x | No | 0.075–0.15 | 12–30 | 29 | Yes | ||
X-Mind Trium “True low dose” | 80–90 | 4–10 | 0.5 | FPD (CMOS) | 12 | x | No | 0.075 | 12–30 | 29 | Yes | |||
X-Mind Prime 3D | 60–86 | 2–12.5 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.08 | 7 | ? | ? | ||
WhiteFox | 105 | 6–10 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.1–0.3 | 18–27 | 30 | Yes | |
Air Techniques (USA) | ProVecta 3D Prime | 50–99 | 4–16 | 0.5 | FPD (CMOS - CsI) | 14 | x | x | x | No | 0.08–0.2 | 2–18 | ? | Yes |
Asahi Roentgen (Japan) | Alphard 2520 | 60–110 | 2–15 | 0.6 | FPD | ? | x | x | No | 0.1–0.39 | ? | ? | ? | |
Alphard 3030 VEGA | 60–110 | 2–15 | 0.6 | FPD | ? | x | x | No | 0.1–0.39 | ? | ? | ? | ||
Alioth | 60–100 | 1–12 | 0.5 | FPD (CsI) | 8–14 | x | x | No | 0.1–0.2 | 17 | ? | No | ||
Alioth CM | 60–100 | 1–12 | 0.5 | FPD (CsI) | 8–14 | x | x | No | 0.1–0.2 | 17 | ? | No | ||
Auge Zio | 60–95 | 2–12 | 0.5 | FPD (CsI) | 12 | x | x | No | 0.1–0.2 | 8.5–17 | ? | No | ||
Auge Zio CM | 60–95 | 2–12 | 0.5 | FPD (CsI) | 12 | x | x | No | 0.1–0.2 | 8.5–17 | ? | No | ||
Auge Zio CM Maxim | 60–95 | 2–12 | 0.5 | FPD (CsI) | 12 | x | x | No | 0.1–0.2 | 8.5–17 | ? | No | ||
Auge X Zio | 60–95 | 2–12 | 0.5 | FPD (CsI) | 12 | x | x | No | 0.1–0.2 | 8.5–17 | ? | No | ||
Auge X Zio CM | 60–95 | 2–12 | 0.5 | FPD (CsI) | 12 | x | x | No | 0.1–0.2 | 8.5–17 | ? | No | ||
Auge X Zio CM Maxim | 60–95 | 2–12 | 0.5 | FPD (CsI) | 12 | x | x | No | 0.1–0.2 | 8.5–17 | ? | No | ||
Auge Solio Z | 60–100 | 2–12 | 0.6 | ? | 12 | x | x | x | No | 0.1–0.3 | ? | ? | ? | |
Auge Solio Z CM | 60–100 | 2–12 | 0.6 | ? | 12 | x | x | x | No | 0.1–0.3 | ? | ? | ? | |
Solio X | 60–85 | 2–8 | 0.5 | ? | ? | x | x | No | 0.1–0.17 | 6–12 | ? | ? | ||
Solio X Z Maxim | 60–85 | 2–8 | 0.5 | ? | ? | x | x | No | 0.1–0.17 | 6–12 | ? | ? | ||
PSR 9000N | ? | ? | ? | ? | ? | ? | ? | ? | ? | 0.1–0.15 | 17 | ? | ? | |
Biolase (USA) | DaVinci Imaging D3D | 90 | 10 | 0.5–0.6 | II (CCD) | 12 | x | x | No | 0.17–0.33 | 10–30 | 30–70 | Yes | |
Carestream (France) | CS 8100 3D | 60–90 | 2–15 | 0.6–0.7 | FPD (CMOS) | 14 | x | x | No | 0.075-? | 7–15 | >120 | Yes | |
CS 8100SC 3D | 60–90 | 2–15 | 0.6–0.7 | FPD (CMOS) | 14 | x | x | No | 0.075-? | 7–15 | >120 | Yes | ||
CS 9000 3D / Kodak 9000 3D | 60–90 | 2–15 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.076-? | 22 | >120 | Yes | ||
CS 9000C 3D / Kodak 9000C 3D | 60–90 | 2–15 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.076-? | 22 | >120 | Yes | ||
CS 9300 | 60–90 | 2–15 | 0.7 | TFT | 14 | x | x | x | No | 0.09–0.5 | 12–28 | >120 | Yes/No | |
CS 9300C | 60–90 | 2–15 | 0.7 | TFT | 14 | x | x | x | No | 0.09–0.5 | 12–28 | >120 | Yes/No | |
CS 9300 Select | 60–90 | 2–15 | 0.7 | TFT | 14 | x | x | No | 0.09–0.5 | 12–28 | >120 | Yes/No | ||
CS 9300C Select | 60–90 | 2–15 | 0.7 | TFT | 14 | x | x | No | 0.09–0.5 | 12–28 | >120 | Yes/No | ||
CS 9300 ENT | 60–90 | 2–15 | 0.7 | TFT | 14 | x | x | x | No | 0.09–0.5 | 12–28 | >120 | Yes/No | |
Kodak 9500 MFOV | 60–90 | 2–15 | 0.7 | FPD (A-Si) | 14 | x | No | 0.2–0.3 | 24 | 80–150 | Yes | |||
Kodak 9500 LFOV | 60–90 | 2–15 | 0.7 | FPD (A-Si) | 14 | x | No | 0.2–0.3 | 24 | 80–150 | Yes | |||
CS 9600 12 × 10 | 60–120 | 2–15 | 0.3–0.7 | FPD (CMOS) | 14 | x | x | x | No | 0.075–0.4 | 5.5–40 | ? | Yes | |
CS 9600 16 × 10 | 60–120 | 2–15 | 0.3–0.7 | FPD (CMOS) | 14 | x | x | x | No | 0.075–0.4 | 5.5–40 | ? | Yes | |
CS 9600 16 × 17 | 60–120 | 2–15 | 0.3–0.7 | FPD (CMOS) | 14 | x | x | x | No | 0.075–0.4 | 5.5–40 | ? | Yes | |
Castellini (Italy) | X Radius Compact 3D | 90 | 4–15 | 0.6 | FPD (A-Si - CsI) | 16 | x | x | No | 0.08–0.16 | ? | ? | Yes | |
X Radius Trio SFOV | 60–90 | 1–10 | 0.5 | FPD (A-Si - CsI) | 16 | x | x | x | Yes | 0.075-? | >18 | >15 | Yes | |
X Radius Trio SFOV (Ceph) | 60–90 | 1–10 | 0.5 | FPD (A-Si - CsI) | 16 | x | x | x | Yes | 0.075-? | >18 | >15 | Yes | |
X Radius Trio LFOV | 60–90 | 1–10 | 0.5 | FPD (A-Si - CsI) | 16 | x | x | x | Yes | 0.075-? | >18 | >15 | Yes | |
X Radius Trio LFOV (Ceph) | 60–90 | 1–10 | 0.5 | FPD (A-Si - CsI) | 16 | x | x | x | Yes | 0.075-? | >18 | >15 | Yes | |
Dürr (Germany) | VistaVOX S | 50–99 | 4–16 | 0.5 | FPD (CMOS - CsI) | ? | x | x | x | No | 0.08–0.12 | 2–18 | ? | ? |
VistaVOX S (Ceph) | 50–99 | 4–16 | 0.5 | FPD (CMOS - CsI) | ? | x | x | x | No | 0.08–0.12 | 2–18 | ? | ? | |
Dabi Atlante (Brazil) | Eagle 3D | 85 | 4–8 | 0.5 | FPD (CMOS) | ? | x | x | x | Yes | 0.08–0.5 | 7–32 | 22–97 | Yes |
Eagle 3D (Ceph) | 85 | 4–8 | 0.5 | FPD (CMOS) | ? | x | x | x | Yes | 0.08–0.5 | 7–32 | 22–97 | Yes | |
Dentium (Republic of Korea) | Rainbow CT | 60–100 | 4–12 | 0.5 | FPD (CMOS) | ? | x | x | Yes | 0.2–0.3 | 19 | >60 | ? | |
Denstply Sirona (Germany) | Orthophos S 3D | 60–90 | 4–13 | 0.5 | FPD (A-Si) | ? | x | x | x | No | 0.08–0.22 | 14 | ? | Yes |
Orthophos S 3D (Ceph) | 60–90 | 4–13 | 0.5 | FPD (A-Si) | ? | x | x | x | No | 0.08–0.22 | 14 | ? | Yes | |
Orthophos SL 3D | 60–90 | 4–13 | 0.5 | FPD (A-Si) | ? | x | x | x | No | 0.08–0.22 | 14 | ? | Yes | |
Orthophos SL 3D (Ceph) | 60–90 | 4–13 | 0.5 | FPD (A-Si) | ? | x | x | x | No | 0.08–0.22 | 14 | ? | Yes | |
Orthophos XG 3D | 60–90 | 4–13 | 0.5 | FPD (A-Si) | ? | x | x | No | 0.1–0.16 | 14 | ? | Yes | ||
Orthophos XG 3D (Ceph) | 60–90 | 4–13 | 0.5 | FPD (A-Si) | ? | x | x | No | 0.1–0.16 | 14 | 90–390 | Yes | ||
Galileos Compact | 85 | 5–7 | 0.5 | II | 12 | x | No | 0.3 | 14 | 240–300 | Yes | |||
Galileos Comfort | 85 | 5–7 | 0.5 | II | 12 | x | No | 0.15–0.3 | 14 | 240–300 | Yes | |||
Galileos Comfort Plus | 98 | 3–6 | 0.5 | II | 12 | x | No | 0.125–0.25 | 14 | ? | ? | |||
Fussen (China) | Dentrix 20 | 60–90 | 4–10 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.075–0.25 | 15 | >60 | Yes | |
FONA Dental (Slovakia) | Stellaris 3D | 60–86 | 2.5–10 | 0.5 | FPD (CMOS) | 14 | x | x | No | 0.08-? | 13–16.9 | ? | Yes | |
Stellaris 3D (Ceph) | 60–86 | 2.5–10 | 0.5 | FPD (CMOS) | 14 | x | x | No | 0.08-? | 13–16.9 | ? | Yes | ||
FONA X PAN 3D | 61–85 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.16 | 12.3 | >10 | Yes | |||
FONA X PAN 3D Plus | 61–85 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.16 | 12.3 | >10 | Yes | |||
Genoray (Republic of Korea) | Volux 6 | 60–85 | 5–7 | 0.5 | ? | ? | x | x | No | 0.1–0.17 | 20 | 180 | ? | |
Volux 9 (Dental/ENT) | 85 | 7 | 0.5 | ? | ? | x | No | ? | 8.5 | 120 | ? | |||
Volux 21 | 60–110 | 5–7 | 0.5 | FPD (CMOS) | ? | x | No | 0.1–0.28 | 15.8 | 150 | ? | |||
Volux 21C | 60–110 | 5–7 | 0.5 | FPD (CMOS) | ? | x | No | 0.1–0.28 | 15.8 | 150 | ? | |||
Papaya 3D | 60–90 | 4–12 | 0.5 | ? | ? | x | x | x | Yes | 0.075–0.4 | 7.7–14.5 | ? | ? | |
Papaya 3D Plus | 60–90 | 4–12 | 0.5 | ? | ? | x | x | x | Yes | 0.075–0.4 | 7.7–14.5 | ? | ? | |
Papaya 3D Premium NV | 60–90 | 4–12 | 0.5 | ? | ? | x | Yes | 0.075–0.4 | >7.7 | ? | ? | |||
Papaya 3D Premium NV (Ceph) | 60–90 | 4–12 | 0.5 | ? | ? | x | Yes | 0.075–0.4 | >7.7 | ? | ? | |||
Papaya 3D Premium LV | 60–90 | 4–12 | 0.5 | ? | ? | x | Yes | 0.075–0.4 | >7.7 | ? | ? | |||
Papaya 3D Premium LV (Ceph) | 60–90 | 4–12 | 0.5 | ? | ? | x | Yes | 0.075–0.4 | >7.7 | ? | ? | |||
Papaya 3D Premium ENT | 60–90 | 4–12 | 0.5 | ? | ? | x | Yes | 0.075–0.4 | >7.7 | ? | ? | |||
HDX Will (Republic of Korea) | Dentri α Classic | 60–110 | 5–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.1–0.3 | 16–36 | 8–40 | Yes | ||
Dentri α Extended | 60–110 | 5–10 | 0.5 | FPD (CMOS) | 14 | x | Yes | 0.1–0.3 | 8–24 | 8–40 | Yes | |||
Dentri Cα Classic | 60–110 | 5–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.1–0.3 | 16–36 | 8–40 | Yes | |||
Dentri Cα Extended | 60–110 | 5–10 | 0.5 | FPD (CMOS) | 14 | x | Yes | 0.1–0.3 | 8–24 | 8–40 | Yes | |||
Dentri Sα Classic | 60–110 | 5–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.1–0.3 | 16–36 | 8–40 | Yes | |||
Dentri Sα Extended | 60–110 | 5–10 | 0.5 | FPD (CMOS) | 14 | x | Yes | 0.1–0.3 | 8–24 | 8–40 | Yes | |||
Dinnova 3 2520D/3030D | 50–120 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.15–0.4 | 7–24 | >120 | Yes | |||
Q-Face (non-stitch) | 60–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.1–0.3 | 8–24 | >60 | Yes | |
Q-Face (one stitch) | 60–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | Yes | 0.1–0.3 | 8–24 | >60 | Yes | |
Q-Face (two stitch) | 60–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | Yes | 0.1–0.3 | 8–24 | >60 | Yes | |
Q-Face-S (non-stitch) | 60–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.1–0.3 | 8–24 | >60 | Yes | |
Q-Face-S (one stitch) | 60–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | Yes | 0.1–0.3 | 8–24 | >60 | Yes | |
Q-Face-S (two stitch) | 60–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | Yes | 0.1–0.3 | 8–24 | >60 | Yes | |
Hitachi (Japan) | CB Mercuray | 60–120 | 10–15 | ? | II (CCD) | 12 | x | x | No | 0.1–0.37 | 9.6 | 360 | ? | |
CB Throne | 60–120 | 10–15 | ? | II (CCD) | 12 | x | x | No | 0.1*2–0.2*4 | 9.6 | ? | ? | ||
ImageWork (USA) | Panoura 18S | 58–82 | 2–10 | 0.5 | ? | 16 | x | x | No | 0.08–0.1 | 11.5–23 | ? | ? | |
Panoura 18S (Ceph) | 58–82 | 2–10 | 0.5 | ? | 16 | x | x | No | 0.08–0.1 | 11.5–23 | ? | ? | ||
J Morita (Japan) | 3D Accuitomo XYZ | 60–80 | 1–10 | 0.5 | II (CCD) | ? | x | No | 0.125 | 18 | >300 | No | ||
3D Accuitomo FPD 60 | 60–80 | 1–10 | 0.5 | FPD (CMOS) | 13 | x | No | 0.08–0.125 | >18 | >180 | No | |||
3D Accuitomo FPD 80 | 60–80 | 1–10 | 0.5 | FPD (CMOS) | 13 | x | x | No | 0.08–0.16 | >18 | >180 | No | ||
3D Accuitomo 170 | 60–90 | 1–10 | 0.5 | FPD (A-Si) | 14 | x | x | x | No | 0.08–0.25 | 5.4–30.8 | >180 | No | |
Veraview X800 F40P | 60–100 | 2–10 | 0.5 | FPD | ? | x | x | No | 0.08–0.125 | 9.4 | <60 | No | ||
Veraview X800 F40PC | 60–100 | 2–10 | 0.5 | FPD | ? | x | x | No | 0.08–0.125 | 9.4 | <60 | No | ||
Veraview X800 R100P | 60–100 | 2–10 | 0.5 | FPD | ? | x | x | No | 0.08–0.125 | 9.4 | <60 | No | ||
Veraview X800 R100PC | 60–100 | 2–10 | 0.5 | FPD | ? | x | x | No | 0.08–0.125 | 9.4 | <60 | No | ||
Veraview X800 F150P | 60–100 | 2–10 | 0.5 | FPD | ? | x | x | x | No | 0.08–0.125 | 9.4 | <60 | No | |
Veraview X800 F150PC | 60–100 | 2–10 | 0.5 | FPD | ? | x | x | x | No | 0.08–0.125 | 9.4 | <60 | No | |
Veraviewpocs 3D F40 | 60–90 | 2–10 | 0.5 | FPD (CMOS) | 13 | x | x | No | 0.125 | 9.4 | <60 | No | ||
Veraviewpocs 3D F40PC | 60–90 | 2–10 | 0.5 | FPD (CMOS) | 13 | x | x | No | 0.125 | 9.4 | <60 | No | ||
Veraviewpocs 3D R100 | 60–90 | 2–10 | 0.5 | FPD (CMOS) | 13 | x | x | No | 0.125–0.16 | 9.4 | <60 | No | ||
Veraviewpocs 3D R100PC | 60–90 | 2–10 | 0.5 | FPD (CMOS) | 13 | x | x | No | 0.125–0.16 | 9.4 | <60 | No | ||
KaVo Kerr (USA/Finland) | KaVo 3D eXam | 120 | 3–7 | 0.5 | FPD (A-Si) | 14 | x | No | 0.2–0.4 | 9–27 | >120 | Yes | ||
KaVo OP 3D | 95 | 2–12.5 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.08–0.4 | 10–20 | ? | ? | |
KaVo OP 3D (Ceph) | 95 | 2–12.5 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.08–0.4 | 10–20 | ? | ? | |
KaVo OP 3D Pro Small Panel | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | ? | x | x | No | 0.085–0.3 | 11–21 | ? | ? | ||
KaVo OP 3D Pro Small Panel (Ceph) | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | ? | x | x | No | 0.085–0.3 | 11–21 | ? | ? | ||
KaVo OP 3D Pro Large Panel | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | ? | x | x | x | Yes | 0.085–0.4 | 11–42 | ? | ? | |
KaVo OP 3D Pro Large Panel (Ceph) | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | ? | x | x | x | Yes | 0.085–0.4 | 11–42 | ? | ? | |
KaVo 3D Vision (V8) | 90–120 | 3–8 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.125–0.4 | 4.8–26.9 | >30 | Yes | ||
KaVo 3D Vision (V10) | 90–120 | 3–8 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.125–0.4 | 4.8–26.9 | >30 | Yes | |
KaVo 3D Vision (V17) | 90–120 | 3–8 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.125–0.4 | 4.8–26.9 | >30 | Yes | |
KaVo (Gendex - USA) | GXDP-800 | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | 16 | x | x | x | No | 0.085–0.4 | 10–20 | ? | Yes |
GXDP-800 (Ceph) | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | 16 | x | x | x | No | 0.085–0.4 | 10–20 | ? | Yes | |
GXDP-700-S | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.085–0.3 | 11–17 | ? | Yes | ||
GXDP-700-SC | 57–90 | 3.2–16 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.085–0.3 | 11–17 | ? | Yes | ||
GXCB-500 | 90–120 | 3–8 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.125–0.4 | 8.9–23 | 20–95 | Yes | ||
KaVo (Imaging Science - USA) | i-CAT Classic | 120 | 3–7 | 0.5 | FPD (A-Si) | 14 | x | No | 0.2–0.4 | 10–40 | >120 | ? | ||
i-CAT Precise | 120 | 3–7 | 0.5 | FPD (A-Si - CsI) | 14 | x | x | x | No | 0.125–0.4 | 4–23 | >95 | ? | |
i-CAT Next Generation | 120 | 3–7 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.125–0.4 | 5–27 | >60 | Yes | ||
i-CAT FLX MV | 120 | 3–8 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.125–0.4 | 4.8–23 | >30 | Yes | |
i-CAT FLX V8 | 120 | 3–8 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.125–0.4 | 4.8–26.9 | >30 | Yes | ||
i-CAT FLX V10 | 120 | 3–8 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.125–0.4 | 4.8–26.9 | >30 | Yes | |
i-CAT FLX V17 | 120 | 3–8 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.125–0.4 | 4.8–26.9 | >30 | Yes | |
KaVo (Instrumentarium – Finland) | OP200 D | 57–85 | 2–16 | 0.5 | ? | ? | x | No | 0.23 | ? | ? | ? | ||
OC 200 D | 57–85 | 2–16 | 0.5 | ? | ? | x | No | 0.23 | ? | ? | ? | |||
OP300 | 57–90 | 4–16 | 0.5 | FPD (CMOS) | 14 | x | x | No | 0.085–0.3 | 11–21 | >120 | Yes | ||
OP300 Maxio | 57–90 | 4–16 | 0.5 | FPD (CMOS) | 14 | x | x | x | Yes | 0.085–0.3 | 11–40 | >120 | Yes | |
KaVo (Soredex – Finland) | Scanora 3D | 60–90 | 8–15 | 0.5 | FPD (CMOS) | 12 | x | x | x | No | 0.133–0.35 | 10–26 | 60–360 | Yes |
Scanora 3DX | 60–90 | 4–10 | 0.5 | FPD (A-Si) | ? | x | x | x | No | 0.1–0.5 | 18–34 | ? | Yes | |
Cranex 3D | 57–90 | 4–16 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.085–0.3 | 10–20 | ? | Yes | ||
Cranex 3D (Ceph) | 57–90 | 4–16 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.085–0.3 | 10–20 | ? | Yes | ||
Cranex 3DX | 57–90 | 4–16 | 0.5 | FPD (CMOS) | 16 | x | x | x | Yes | 0.085–0.4 | 10–40 | ? | Yes | |
Cranex 3DX (Ceph) | 57–90 | 4–16 | 0.5 | FPD (CMOS) | 16 | x | x | x | Yes | 0.085–0.4 | 10–40 | ? | Yes | |
LargeV (China) | HiRes 3D Dental | 60–100 | 2–10 | 0.5 | FPD (CMOS) | ? | x | x | x | Yes | 0.075–0.25 | 15–21 | >30 | Yes |
HiRes 3D MAX | 60–100 | 2–10 | 0.5 | FPD (A-Si) | ? | x | x | Yes | 0.075–0.3 | 15 | >40 | Yes | ||
Smart Dental | 60–100 | 2–10 | 0.4 | FPD (CMOS) | ? | x | x | x | No | 0.05–0.25 | 13 | 20–40 | ? | |
Smart Dental (Ceph) | 60–100 | 2–10 | 0.4 | FPD (CMOS) | ? | x | x | x | No | 0.05–0.25 | 13 | 20–40 | ? | |
MyRay (Italy) | SkyView | 90 | 10 (max) | 0.5–0.6 | II | 12 | x | x | No | 0.17–0.33 | 10–30 | 30–70 | Yes | |
Hyperion X5 3D/2D | 90 | 4–15 | 0.6 | FPD (A-Si) | 16 | x | x | No | 0.08–0.16 | 6.4–16.8 | >15 | Yes | ||
Hyperion X5 3D/2D CEPH | 90 | 4–15 | 0.6 | FPD (A-Si) | 16 | x | x | No | 0.08–0.16 | 6.4–16.8 | >15 | Yes | ||
Hyperion X9 Full FOV | 60–90 | 1–10 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075-? | 18 | >15 | Yes | ||
Hyperion X9 Full FOV CEPH | 60–90 | 1–10 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075-? | 18 | >15 | Yes | ||
Hyperion X9 Extended FOV | 60–90 | 1–10 | 0.5 | FPD (A-Si) | 16 | x | x | x | Yes | 0.075-? | 18 | >15 | Yes | |
Hyperion X9 Extended FOV CEPH | 60–90 | 1–10 | 0.5 | FPD (A-Si) | 16 | x | x | x | Yes | 0.075-? | 18 | >15 | Yes | |
Hyperion X9 Pro 10 × 8 version | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075-? | 18 | >15 | Yes | ||
Hyperion X9 Pro CEPH 10 × 8 version | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075-? | 18 | >15 | Yes | ||
Hyperion X9 Pro 13 × 16 version | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.075-? | 18 | >15 | Yes | |
Hyperion X9 Pro CEPH 13 × 16 version | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.075-? | 18 | >15 | Yes | |
Meyer (China) | SS-91010D Pro-3D | 50–90 | 2–10 | 0.5 | FPD (CMOS) | ? | ? | ? | ? | ? | ? | 11.7–20 | ? | ? |
SS-91010D Pro-3De | 50–90 | 2–10 | 0.5 | FPD (CMOS) | ? | ? | ? | ? | ? | ? | 11.7–20 | ? | ? | |
X12008D Pro-3D | 60–120 | 2–8 | ? | ? | ? | x | ? | 0.24 | 20 | 60 | ? | |||
PiXAMED (Thailand) | DentiiScan 2.0 | 90 | 6–9 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.2–0.4 | 18 | 8–18 | Yes | |
Owandy (France) | I-MAX 3D | 60–86 | 2–12.5 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.087–0.175 | 10.8–11.2 | ? | ? |
I-MAX 3D Touch | 60–86 | 6–10 | 0.5 | FPD (A-Si) | ? | x | x | No | 0.092 | 20 | >90 | Yes | ||
I-MAX 3D Touch CEPH | 60–86 | 6–10 | 0.5 | FPD (A-Si) | ? | x | x | No | 0.092 | 20 | >90 | Yes | ||
Osstem (Republic of Korea) | Osstem Implant CBCT T1 | 60–100 | 5–16 | 0.5 | FPD | ? | x | x | No | 0.2 | 14–22 | 40 | ? | |
Planmeca (Finland) | Promax 3D Classic | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 12 | x | x | x | Yes | 0.075–0.4 | 9–37 | 2–25 | Yes |
Promax 3D Classic (Ceph) | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 12 | x | x | x | Yes | 0.075–0.4 | 9–37 | 2–25 | Yes | |
Promax 3D S | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | Yes | 0.075–0.4 | 7.5–27 | 2–25 | Yes | ||
Promax 3D S (Ceph) | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | Yes | 0.075–0.4 | 7.5–27 | 2–25 | Yes | ||
Promax 3D Plus | 60–120 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | No | 0.075–0.6 | 9–33 | 2–30 | Yes | ||
Promax 3D Plus (Ceph) | 60–120 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | No | 0.075–0.6 | 9–33 | 2–30 | Yes | ||
Promax 3D Plus ENT | 60–120 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | x | No | 0.075–0.6 | 9–33 | 2–30 | Yes | |
Promax 3D Plus ENT (Ceph) | 60–120 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | x | No | 0.075–0.6 | 9–33 | 2–30 | Yes | |
Promax 3D Mid | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | x | Yes | 0.075–0.6 | 9–33 | 2–55 | Yes | |
Promax 3D Mid (Ceph) | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | x | Yes | 0.075–0.6 | 9–33 | 2–55 | Yes | |
Promax 3D Mid ENT | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | x | Yes | 0.075–0.6 | 9–33 | 2–55 | Yes | |
Promax 3D Mid ENT (Ceph) | 54–90 | 1–14 | 0.5 | FPD (CMOS) | 15 | x | x | x | Yes | 0.075–0.6 | 9–33 | 2–55 | Yes | |
Promax 3D Max | 54–96 | 1–12.5 | 0.5–0.6 | FPD (CMOS) | 15 | x | x | x | Yes | 0.075–0.4 | 9–55 | 2–55 | Yes | |
Promax 3D Max ENT | 54–96 | 1–12.5 | 0.5–0.6 | FPD (CMOS) | 15 | x | x | x | Yes | 0.075–0.6 | 9–55 | 2–55 | Yes | |
Viso | 80–120 | 1–16 | 0.5 | FPD | ? | x | x | x | No | 0.075–0.6 | 1–36 | 2–55 | Yes | |
Viso (Ceph) | 80–120 | 1–16 | 0.5 | FPD | ? | x | x | x | No | 0.075–0.6 | 1–36 | 2–55 | Yes | |
Viso (Vertical stitching) | 80–120 | 1–16 | 0.5 | FPD | ? | x | x | x | Yes | 0.075–0.6 | 1–36 | 2–55 | Yes | |
Viso (Vertical stitching) (Ceph) | 80–120 | 1–16 | 0.5 | FPD | ? | x | x | x | Yes | 0.075–0.6 | 1–36 | 2–55 | Yes | |
Pointnix (Republic of Korea) | Point I3D | 50–90 | 4–16 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.23–0.47 | 19 | 19–24 | ? | |
Point 3D Combi 500 | 50–90 | 4–10 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.18–0.43 | 19 | 10–40 | Yes | ||
Point 3D Combi 500c | 50–90 | 4–10 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.18–0.43 | 19 | 10–40 | Yes | ||
Point 3D Combi 500 s | 50–90 | 4–10 | 0.5 | FPD (CMOS - CsI) | 14 | x | x | No | 0.18–0.43 | 19 | 10–40 | Yes | ||
Point 800 HD Plus | 50–90 | 4–10 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.18–0.43 | 19 | 10–40 | ? | ||
Point 800c HD Plus | 50–90 | 4–10 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.18–0.43 | 19 | 10–40 | ? | ||
Point 800 s HD Plus | 50–90 | 4–10 | 0.5 | ? | 14 | x | x | No | 0.18–0.43 | 19 | 10–40 | ? | ||
PreXion (Japan) | PreXion 3D | 90 | 4 | 0.15 | FPD (CsI) | 16 | x | No | 0.1–0.15 | 19–37 | >60 | No | ||
PreXion 3D Elite | 90 | 4 | 0.2 | FPD (CsI) | 14 | x | x | No | 0.1–0.16 | 8.6–33.5 | 30 | No | ||
PreXion 3D Elite Element | 90 | 4 | 0.2 | FPD (CsI) | 14 | x | No | 0.1–0.16 | 8.6–33.5 | 30 | No | |||
PreXion 3D Eclipse | 90 | 4 | 0.2 | FPD (CsI) | 14 | x | No | 0.15 | 8.7–17.4 | >30 | No | |||
PreXion 3D Eclipse (Ceph) | 90 | 4 | 0.2 | FPD (CsI) | 14 | x | No | 0.15 | 8.7–17.4 | >30 | No | |||
PreXion 3D Excelsior Ex | 60–110 | 1–6 | 0.3 | FPD (CsI) | 16 | x | x | No | 0.08–0.2 | 5.2–23 | ? | No | ||
PreXion 3D Excelsior Ex (Ceph) | 60–110 | 1–6 | 0.3 | FPD (CsI) | 16 | x | x | No | 0.08–0.2 | 5.2–23 | ? | No | ||
PreXion 3D Excelsior Plus | 60–110 | 1–6 | 0.3 | FPD (CsI) | 16 | x | x | No | 0.08–0.2 | 5.2–23 | ? | No | ||
PreXion 3D Excelsior Plus (Ceph) | 60–110 | 1–6 | 0.3 | FPD (CsI) | 16 | x | x | No | 0.08–0.2 | 5.2–23 | ? | No | ||
PreXion 3D Excelsior Pro | 60–110 | 1–6 | 0.3 | FPD (CsI) | 16 | x | x | x | No | 0.08–0.2 | 5.2–23 | ? | No | |
PreXion 3D Excelsior Pro (Ceph) | 60–110 | 1–6 | 0.3 | FPD (CsI) | 16 | x | x | x | No | 0.08–0.2 | 5.2–23 | ? | No | |
PreXion 3D Explorer | 90–110 | 1–5 | 0.3 | FPD (A-Si TFT) | 16 | x | x | x | No | 0.074–0.3 | 10–20 | 60 | Yes | |
Quantitative Radiology / Cefla Dental Group (Italy) | Giano HR Prime | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075–0.3 | 3.6–26.4 | >15 | Yes | |
Giano HR Prime (Ceph) | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075–0.3 | 3.6–26.4 | >15 | Yes | ||
Giano HR Advanced | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.068–0.3 | 6.4–33.6 | >60 | Yes | |
Giano HR Advanced (Ceph) | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.068–0.3 | 6.4–33.6 | >60 | Yes | |
Giano HR Professional | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.068–0.3 | 6.4–33.6 | >60 | Yes | |
Giano HR Professional (Ceph) | 90 | 2–16 | 0.5 | FPD (A-Si) | 16 | x | x | x | No | 0.068–0.3 | 6.4–33.6 | >60 | Yes | |
Giano/NewTom VG3 | 60–90 | 1–10 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075-? | 18 | >15 | Yes | ||
Giano/NewTom VG3 (Ceph) | 60–90 | 1–10 | 0.5 | FPD (A-Si) | 16 | x | x | No | 0.075-? | 18 | >15 | Yes | ||
Go 2D/3D Imaging | 90 | 4–15 | 0.6 | FPD (A-Si) | 16 | x | x | No | 0.08-? | 6.4–16.8 | >15 | Yes | ||
NewTom 9000 Maxiscan | 110 | 1–15 | 1.2 × 0.8 | ? | ? | x | No | 0.3–1.2 | 70 | ? | Yes | |||
NewTom VG | 110 | 1–20 | 0.3 | FPD (A-Si) | 14 | x | No | 0.3 | 24 | 180 | Yes | |||
NewTom VG Flex | 110 | 1–20 | 0.3 | FPD (A-Si) | 14 | No | 0.3 | 24 | 180 | Yes | ||||
NewTom VGi | 110 | 1–20 | 0.3 | FPD (A-Si) | 16 | x | x | x | No | 0.075–0.3 | 18–26 | 60 | Yes | |
NewTom VGi Flex | 110 | 1–20 | 0.3 | FPD (A-Si) | 14 | x | x | x | No | 0.075–0.3 | 18–26 | 60 | Yes | |
NewTom VGi evo | 75–110 | 1–32 | 0.3 | FPD (A-Si) | 16 | x | x | x | No | 0.1–0.3 | 15–25 | 60 | Yes | |
NewTom 3G | 110 | 15 | 0.5 | II (CCD) | 12 | x | x | No | 0.3 | 36 | 300 | Yes | ||
NewTom 5G | 110 | 1–20 | 0.3 | FPD (A-Si) | 16 | x | x | x | No | 0.075–0.3 | 18–36 | 300 | Yes | |
NewTom 5G XL | 110 | 1–20 | 0.3 | FPD (A-Si) | 16 | x | x | x | No | 0.075–0.3 | 18–36 | 300 | Yes | |
Ritter Imaging (Germany) | Orion | ? | ? | ? | ? | ? | ? | ? | ? | ? | 0.1–0.167 | ? | ? | ? |
Ray Medical (Republic of Korea) | RAYSCAN Symphony BC | 60–90 | 4–10 | 0.5 | FPD | 16 | x | No | 0.14 | 20–40 | ? | ? | ||
RAYSCAN Symphony V | 60–90 | 4–10 | 0.5 | FPD | 16 | x | No | 0.19–0.38 | 20–40 | ? | ? | |||
RAYSCAN α−3D | 60–90 | 4–17 | 0.5 | TFT | ? | x | No | 0.16–0.3 | 4.9–14 | ? | Yes | |||
RAYSCAN α−3D (SC, OCS, OCL) | 60–90 | 4–17 | 0.5 | TFT | ? | x | No | 0.16–0.3 | 4.9–14 | ? | Yes | |||
RAYSCAN α + 130 | 60–90 | 4–17 | 0.5 | TFT | ? | x | x | x | No | 0.07–0.3 | 4.9–14 | ? | Yes | |
RAYSCAN α + 130 (SC, OCS, OCL) | 60–90 | 4–17 | 0.5 | TFT | ? | x | x | x | No | 0.07–0.3 | 4.9–14 | ? | Yes | |
RAYSCAN α + 160 | 60–90 | 4–17 | 0.5 | TFT | ? | x | x | x | No | 0.07–0.3 | 4.9–14 | ? | Yes | |
RAYSCAN α + 160 (SC, OCS, OCL) | 60–90 | 4–17 | 0.5 | TFT | ? | x | x | x | No | 0.07–0.3 | 4.9–14 | ? | Yes | |
RAYSCAN m+ | 60–90 | 4–17 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.18–0.4 | ? | ? | Yes | |
RAYSCAN m + SC | 60–90 | 4–17 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.18–0.4 | ? | ? | Yes | |
RAYSCAN m + DR (with chest X-ray) | 60–90 | 4–17 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.18–0.4 | ? | ? | Yes | |
Ray Medical (Republic of Korea) / Apteryx (USA) | RAYSCAN α-Edge | 60–90 | 4–17 | 0.5 | FPD (CMOS) | 16 | x | No | 0.14–0.29 | 14 | ? | Yes | ||
RAYSCAN α-SM Edge | 60–90 | 4–17 | 0.5 | FPD (CMOS) | 16 | x | No | 0.14–0.29 | 14 | ? | Yes | |||
Streamhealth Dental (USA) | Trophypan Smart 3D | 60–90 | 2–15 | 0.7 | FPD (CMOS) | 14 | x | x | No | 0.075–0.4 | 7–15 | ? | ? | |
Suni Medical Imaging (USA) | Suni3D | 40–99 | 2–10 | 0.5 | ? | ? | x | No | 0.08–0.2 | 15–24 | >60 | Yes | ||
Suni3D (Ceph) | 40–99 | 2–10 | 0.5 | ? | ? | x | No | 0.08–0.2 | 15–24 | >60 | Yes | |||
Suni Q3D | 60–90 | 4–12 | 0.5 | FPD (CMOS) | ? | x | x | x | Yes | 0.075–0.4 | 7.7–14.5 | ? | ? | |
Suni Q3D (Ceph) | 60–90 | 4–12 | 0.5 | FPD (CMOS) | ? | x | x | x | Yes | 0.075–0.4 | 7.7–14.5 | ? | ? | |
Trident (Italy) | X-View Cone Beam | 61–85 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.16 | 15 | >10 | Yes | ||
X-View Cone Beam (Ceph) | 61–85 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | No | 0.16 | 15 | >10 | Yes | |||
Vatech (Republic of Korea) | PaX i3D | 50–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | No | 0.12–0.3 | 15–24 | ? | ? | |
PaX i3D (SC or OP) | 50–90 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | No | 0.12–0.3 | 15–24 | ? | ? | ||
PaX i3D Smart / i3D Smart 8 × 8 | 60–99 | 4–16 | 0.5 | ? | 14 | x | No | 0.2–0.3 | 13.5 | 95 | ? | |||
PaX i3D Smart (Ceph)/ i3D Smart 8 × 8 SC | 60–99 | 4–16 | 0.5 | ? | 14 | x | No | 0.2–0.3 | 13.5 | 95 | ? | |||
PaX i3D Smart 2/Green Smart / Smart Plus / i3D Smart | 60–99 | 4–16 | 0.5 | ? | 14 | x | x | x | No | 0.08–0.3 | 18 | ? | ? | |
PaX i3D Smart 2 (Ceph) / Green Smart SC / Smart Plus RC / i3D Smart RC | 60–99 | 4–16 | 0.5 | ? | 14 | x | x | x | No | 0.08–0.3 | 18 | ? | ? | |
PaX i3D Green / Green CT | 50–100 | 4–16 | 0.5 | ? | 14 | x | x | x | No | 0.08–0.3 | 9–15 | ? | ? | |
PaX i3D Green SC or OP / Green CT SC or OP | 50–100 | 4–16 | 0.5 | ? | 14 | x | x | x | No | 0.08–0.3 | 9–15 | ? | ? | |
PaX i3D 2/Green CT 2/Green 16 | 60–99 | 4–16 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.08–0.3 | 4.9–9 | ? | ? | |
PaX i3D 2 (Ceph) / Green CT 2 (Ceph) / Green 16 (Ceph) | 60–99 | 4–16 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.08–0.3 | 4.9–9 | ? | ? | |
Green 18 | 60–99 | 4–16 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.08–0.3 | 4.9–9 | ? | ? | |
Green 18 (Ceph) | 60–99 | 4–16 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.08–0.3 | 4.9–9 | ? | ? | |
Green 21/i3D Premium (AutoCeph) | 60–120 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | No | 0.2–0.4 | 18 | ? | ? | ||
PaX Flex 3D P | 50–90 | 2–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.12–0.3 | 15–24 | >90 | Yes | |
PaX Flex 3D PC | 50–90 | 2–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.12–0.3 | 15–24 | >90 | Yes | |
PaX Uni 3D P | 50–90 | 4–10 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.12–0.3 | 15–24 | >90 | Yes | |
PaX Uni 3D PC | 50–90 | 4–10 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.12–0.3 | 15–24 | >90 | Yes | |
PaX Duo 3D | 50–90 | 2–10 | 0.5 | FPD (CMOS) | ? | x | x | x | No | 0.12–0.3 | 15–24 | >90 | Yes | |
PaX Zenith 3D | 50–120 | 4–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.08–0.3 | 15–24 | >220 | Yes | |
PaX Reve 3D | 50–90 | 2–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.08–0.25 | 15–24 | >120 | ? | |
PaX Reve 3D OS | 50–90 | 2–10 | 0.5 | FPD (CMOS) | 14 | x | x | x | No | 0.08–0.25 | 15–24 | >120 | ? | |
Picasso Trio | 50–90 | 2–10 | 0.5 | FPD (CMOS) | 12 | x | x | x | No | 0.2 | 15–24 | 29 | ? | |
Picasso Trio (Ceph) | 50–90 | 2–10 | 0.5 | FPD (CMOS) | 12 | x | x | x | No | 0.2 | 15–24 | 29 | ? | |
Picasso Duo 3D | 60–90 | 2–10 | 0.5 | FPD | 16 | x | No | ? | 18–24 | 20 | ? | |||
Picasso Master 3D | 40–90 | 2–10 | 0.5 | FPD | 14 | x | No | 0.16–0.2 | 24 | 29 | ? | |||
Picasso Pro 3D | 40–90 | 2–10 | 0.5 | ? | ? | x | No | 0.2–0.3 | 15 | 15 | ? | |||
PaX 500 ECT | 40–90 | 2–10 | 0.5 | FPD | 14 | x | No | 0.2 | 8 | 90 | ? | |||
PaX 500 ECT (Ceph) | 40–90 | 2–10 | 0.5 | FPD | 14 | x | No | 0.2 | 8 | 90 | ? | |||
Villa Sistemi Medicalli (Italy) | Rotograph Evo 3D | 60–86 | 6–10 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.16 | 18–20 | 18–40 | ? | |
Rotograph Evo 3D (Ceph) | 60–86 | 6–10 | 0.5 | FPD (A-Si) | 14 | x | x | No | 0.16 | 18–20 | 18–40 | ? | ||
Rotograph Prime 3D | 60–86 | 2–12.5 | 0.5 | FPD (CMOS) | 16 | x | x | No | 0.087–0.17 | 6.2–9 | ? | ? | ||
UEG Medical (China) | Spectral | 80–110 | 4–8 | 0.5 | FPD (A-Si) | 16 | x | No | 0.075–0.3 | ? | >5 | Yes | ||
Xoran Technologies (USA) | Minicat | 120–125 | 7 | 0.3 | ? | 8 | ? | ? | x | ? | 0.2–0.44 | 10–30 | ? | ? |
Minicat IQ (ENT) | 120 | 7 | ? | ? | ? | ? | ? | ? | ? | ? | 10–30 | ? | ? | |
Minicat 2020 | 120 | 5.8 | ? | ? | ? | x | No | ? | ? | ? | ? | |||
XCat (Portable) | 120 | 6 | 0.15 | FPD (CsI) | 8 | x | No | 0.4 | 20 | 30–120 | ? | |||
Yoshida (Japan) | Finecube | 90 | 4 | 0.15 | FPD (CsI) | 14 | x | x | No | 0.1–0.16 | 8.6–34 | >60 | No | |
X-era Smart F+ | 60–82 | 2–10 | ? | FPD (CMOS) | 16 | x | No | 0.08–0.1 | 11.5–23 | ? | ? | |||
X-era Smart F+ (Ceph) | 60–82 | 2–10 | ? | FPD (CMOS) | 16 | x | No | 0.08–0.1 | 11.5–23 | ? | ? | |||
Yoshida (Japan) / ImageWorks (USA) | X-era NF / Panoura X-era NF | 70–90 | 2–4 | 0.2 | FPD (CMOS) | 16 | x | x | Yes | 0.09–0.15 | 12–20 | ? | ? | |
X-era NF (Ceph) / Panoura X-era NF (Ceph) | 70–90 | 2–4 | 0.2 | FPD (CMOS) | 16 | x | x | Yes | 0.09–0.15 | 12–20 | ? | ? | ||
X-era MF / Panoura X-era MF | 70–90 | 2–4 | 0.2 | FPD (CMOS) | 16 | x | x | Yes | 0.09–0.23 | 12–20 | ? | ? | ||
X-era MF (Ceph) / X-era MF (Ceph) | 70–90 | 2–4 | 0.2 | FPD (CMOS) | 16 | x | x | Yes | 0.09–0.23 | 12–20 | ? | ? |
CBCT, cone beam CT; 2D, two-dimensional; 3D, three-dimensional; FOV, field of view; kV: kilovoltage; mA: miliamperes (tube current).‘CBCT mode’, the lower end of the reported kV range may only apply to panoramic and cephalometric modes.
Further information regarding detector components (e.g. scintillator) is shown in accordance with the terminology used by the manufacturer.
The FOV was divided according to the irradiated surface considering FOV height multiplied by its diameter. A small FOV was considered when the surface was ≤40 cm2; a medium FOV was considered if the surface was >40 cm2 and ≤100 cm2; and a large FOV if it was >100 cm2.
Scan time according to the values reported by the manufacture, although some of these values may actually refer to exposure time but no clear distinction could be made.
All CBCT systems use 2D-array detectors to generate and record projection images. Older technology devices have an image intensifier (II) along with a charge coupled device (CCD) that results in a spherical FOV and is larger and bulkier. Nowadays, the majority CBCT devices use flat panel detectors (FPD). II detectors were used in 3% of the units (only one unit is still available on the market), while 13% of the CBCT machines did not have information regarding the detector type available (Table 2). Nearly one-third (30%) of the manufactures did not report the bit depth of the detectors. From the reported CBCT devices, the bit depth ranged between 8 and 16 bits, with 14 bits being the most commonly used bit depth in all devices (Table 2).
Distribution of FOV options among the CBCT devices is displayed in Figure 4. Of all the CBCT devices examined, the smallest FOV was 20 × 20 mm (X1, 3Shape, Denmark). The device with the largest FOV was Viso (Planmeca, Finland), in which it is possible to achieve a 300 × 300 mm FOV via vertical stitching. A minority of the CBCT devices were equipped with Small/Large (n = 6), Small only (n = 14), Medium only (n = 16) and Medium/Large (n = 16) FOV options. A selection of FOVs comprising all three categories were found for 102 devices, and Small/Medium FOVs for 79 devices. The use of stitching method to extend FOV was clearly reported by the manufactures in 48 CBCT models.
Voxel sizes varied between 0.05 mm (Smart Dental, LargeV, China) and 0.6 mm (Planmeca devices). All the machines presented isotropic voxels, except for one outdated machine (CB Throne, Hitachi, Japan) with a reported anisotropic voxel size of 0.1 × 2 mm or 0.2 × 4 mm.24 Scanning time of devices ranged between 1 and 55 s, although some of the reported scanning times are more likely to be related to the exposure time, especially for those presenting pulsed beam. Devices with the smallest scan times are Orthophos SL 3D, Orthophos S 3D, Orthophos XG 3D and Galileos Comfort Plus with a range of 2–5 s according to the manufacturer. Reconstruction time may vary depending on the workstation, but it is reported to be between 10 and 390 s (Table 2).
Information regarding the X-rays beam operation (i.e. pulsed or continuous) was not found for 32% of the devices. Of the remaining devices, the X-rays beam generation is pulsed in 78% of the machines, and 19% devices used continuous exposure. Five CBCT models (CS9300 Series, Carestream, France) reported both pulsed and not pulsed X-rays beam generation depending on the scanning mode (Table 2).
Considering the wide variation in reported technical characteristics and clinical performance of the available dentomaxillofacial CBCT devices, it is advised to aim for structured feature reporting, thus allowing identification of the proper devices fitting the clinical needs and encompassing the research questions. The suggested standardisation is shown in Table 3.
Table 3.
Features | Specific features | Descriptiona |
---|---|---|
General | Dimensions | Width x depth x (maximum) height in meters |
Weight | Total machine weight in kilograms, without workstation | |
Patient positioning | Standing, sitting, supine. Wheelchair accessibility | |
Software (acquisition) | Software used for image acquisition | |
Software (DICOM viewer) | Software used as DICOM viewer Availability of DICOM structured report tools |
|
X-rays | Tube voltage (kV) | Fixed (exact value) or variable (range) |
Tube current (mA) | Fixed (exact value) or variable (range) | |
Focal spot size | Focal spot size in millimeters | |
X-rays beam generation | Pulsed or continuous x-rays generation | |
Detector | Detector type | Type of the detector used for CBCT acquisitions (incl. scintillator and signal read-out system) |
Detector dimensions | Detector width x height in centimeters | |
Detector resolution | Detector width x height in pixels Binning (if applicable) |
|
Detector bit depth | Signal range of the detector | |
Volume and resolution | Available FOV | All available possibilities of FOV in the machine, not just the smallest and the biggest |
Stitching for extended FOV | The use of stitching to achieve bigger volumes. Specify if it is horizontal or vertical (or both) stitching | |
Voxel size | All available voxel sizes, not just the smallest and the biggest | |
Rotation angle | Rotation arc of the gantry during acquisition | |
Times | Scan time | Time taken for the whole scan (mean and range) |
Exposure time | Time taken only when the X-rays generation is occurring (mean and range) | |
Reconstruction time | Time taken for the volume reconstruction (mean and range) | |
Radiation dose | Dose index | Range of RDSR-compliant dose index values (preferably dose-area product, possibly dose-length product), considering low- and high-dose protocols |
Effective dose | Range of effective dose (in µSv) calculated in an appropriate phantom, considering low- and high-dose protocols |
CBCT, cone beam CT; FOV: field of view; RDSR: radiation dose structured report; µSv: micro-sieverts;kV: kilovoltage; mA: miliamper.
Importantly, for hybrid machines, the CBCT features should be reported separately from those of cephalometric and panoramic imaging.
Discussion
Dental X-ray examination and diagnosis have made great progress over the last 30 years. The number of manufacturers and devices using CBCT technology is growing rapidly. The first CBCT device (NewTom 9000 Maxiscan; Quantitative Radiology/Cefla Dental Group, Italy) was described in 1998 by Mozzo et al.9 Since then, there has been an increasing and continuous development of CBCT machines incorporating new technologies. In the present study, 279 CBCT models (143 CBCT series) from 47 manufacturers were catalogued. Only a few studies explored the features of several CBCT devices, yet there are no recent studies comparing those features for all machines. De Vos et al,20 Kau et al,18 and Nemtoi et al1 conducted similar reviews about the features of CBCT devices available before 2013.
An important development of second-generation CBCT devices was the detector type. Nowadays most of the devices use FPDs, which are distortion-free and show a wider dynamic range compared with the image intensifier detectors used in earlier CBCT models.25 The bit depth of detectors in most current CBCT devices is between 14 and 16 bits and seems reasonable for dentomaxillofacial practice.
Crucial differences among the CBCT devices were found in terms of machine size and weight, available imaging modalities, exposure parameters (kV and mA) and exposure mode and time, FOV size, voxel size, and both scanning and reconstruction time. Machine size and weight is important considering the often-limited physical space available to install a CBCT. One may also consider the imaging modalities required for clinical practice, as most of the machines are either 2-in-1 or 3-in-1. This choice also has an impact in the required room space for installing a machine. From what was observed in the previous compilation of CBCT,1 there has been an upward trend in developing hybrid devices, having the potential for both 2D and 3D extra oral imaging. Such devices are usually applying a smaller detector size, allowing a lower cost and a compact device, for upright patient imaging.
In contrast, the use of exclusive high-end CBCT devices is rapidly decreasing for the same reasons, which is unfortunate considering the often superior image quality of such dedicated devices.
X-ray beam quality and flux are determined by several parameters, such as tube voltage (kV) and tube current (mA).12 These parameters may be fixed for a given CBCT unit, or they can vary between pre-set exposure programs (e.g. large/small, adult/paediatric pre-sets) or manually adjusted by the operator.12,17 In the present overview, it was found that most of the current devices have adjustable kV and mA options. One must bear in mind that exposure parameters affect not only image quality, but also patient radiation dose, and therefore they should be adjusted to optimise the CBCT acquisition, maintaining image quality for diagnosis (as low as reasonably achievable – ALARA; and as low as diagnostically acceptable being indication-oriented and patient-specific– ALADAIP principles).26–28
Apart from exposure parameters, FOV selection also is crucial when acquiring a CBCT scan, as its size is related to the radiation dose.3,11,17 Furthermore, X-ray scatter is notably increased for large FOVs, which can impair image quality.29 It is also not practical to reconstruct larger FOVs at small voxel sizes due to excessive reconstruction time and file size; the use of a larger voxel size limits the spatial resolution,30 as discussed further below. FOVs can be divided into three categories: small, medium and large,25,31 with various combinations of diameter and height within these categories. FOVs may cover a few teeth, an entire jaw, or even the entire skull of the patient. It is advantageous to have devices with a selectable FOV so that the radiation dose given to the patient can be decreased depending on the indication of the examination (e.g. endodontic evaluation vs orthognathic surgery planning),11,32 although small and medium FOVs may cover most dental applications.25 To acquire bigger FOVs, CBCT device should have a large detector size, compatible to the FOV dimensions. However, a larger detector means more expensive machines. To overcome this additional expense in the manufacture, some devices employ automatic stitching of multiple scan volumes, either horizontal or vertical, to acquire larger FOVs.12 In some of the machines, an option to extend the FOV is available as an upgradable feature. CBCT images acquired using FOV stitching imply the acquisition of two (or more) adjacent exams with common areas to allow the fusion of the volumes into one, and therefore double radiation exposure in the overlap region.12 From a clinical point of view, it seems to interfere on image quality and artefacts expression, but this relation should be further investigated.
Usually, the FOV and voxel size are connected as for small FOV, smaller voxel sizes can be selected.3 A wide range of voxel sizes was found in the present study, ranging from 0.05 to 0.6 mm, and they should be selected according to the diagnostic task.3 Smaller voxel dimensions may be associated with better spatial resolution and may thus be necessary when a high level of detail is required. On the other hand, for a given FOV, a smaller voxel size is associated with a higher mAs, thus increasing the radiation dose to the patient. In addition, noise is increased.3,12,32 Therefore, CBCT devices should offer a number of different voxel sizes, so that the examinations allow the use of the largest voxel size (i.e. lowest radiation dose) while maintaining acceptable diagnostic accuracy.
The resolution of a CBCT image is also influenced by other parameters such as the focal spot size, number of projections (rotation angle & frame rate), reconstruction algorithm, scatter and patient motion.3,32 Patient positioning during image acquisition may affect the probability of motion artefacts. Patients in a standing position, as in the great majority of the devices, can be more susceptible to motion, especially in cases with motion disabilities.3 However, no studies have evaluated how the patient positioning can influence motion artefacts.33 Additionally, the scan time may play a role. In the present study, the mean scan time was between 17.5 s, depending on the FOV and voxel size selection. Shortening the scan time could contribute to avoid (but not eliminate) motion artefacts considering that the patient could remain still during a fast scan.33 However, if the patient moves during a shorter scan, the resulting motion artefact may be more pronounced than in a longer scan. Importantly, manufacturers should correctly report the scan time and the exposure time. The scan time correspond to the time between the first and last basis image acquisition while the exposure time is only related to those moments when the patient is exposed to radiation. In some machines very low scan time were reported (e.g. 1 s) and it is more likely to be the exposure time. The scanning time is only equal to the exposure time for CBCT devices that present a continuous beam operation mode (i.e. not pulsed). However, most of the CBCT devices presented a pulsed beam mode; therefore the scanning time is higher than the exposure time, but with the advantage of reducing radiation dose to the patient.3
Information on technical aspects of the machines are usually accessible on the company’s official website and downloadable brochures and device manuals. However, on some websites this information is limited. Even in brochures and manuals, one may not find all the technical information about the CBCT device. This difficulty was previously reported,1 thus some information could not be displayed in our current overview. This also applies to effective radiation dose range of the CBCT devices, which was probably the least reported feature (less than 40% of the machines). The lack of standardised data and the missing scientific reports comparatively assessing dose levels, prevented us from reporting tabulated comparative information on dosimetry. In a further report, an attempt will be made to compare the scientific output data for patient-specific and indication-oriented radiation dose levels. From the available evidence, we can summarise that differences in radiation dose levels are huge, both between and within specific CBCT devices. While some devices can provide 3D image data at the dose level of two to five panoramic images, such or other CBCT devices can also produces dose levels as high as medical CT when orienting towards another indication and parameter set-up. A 50-fold radiation dose difference can be easily seen when changing the settings in specific CBCT devices.27,34 A standardisation of the technical aspects and features reported regarding the CBCT devices would benefit researchers and practitioners when considering acquiring a machine (see also Table 3).
Conclusion
In conclusion, 279 CBCT devices models were catalogued. Given the wide range of CBCT devices available on the market at the time of writing, dental CBCT should be considered as a generic name applicable to a group of heterogeneous devices. The variability of CBCT features makes it impossible to draw general comparisons between different models, especially in the research field, since CBCT devices may present more variables related to the final image then the tested parameter. Therefore, a systematic review should be carried out to compare published research evidence on indications, image quality and radiation dose levels of all CBCTs on the market. The information tabulated in the present study will be later provided on the International Association of DentoMaxilloFacial Radiology website (www.iadmfr.one). Considering ongoing developments and continued improvements, the present overview will need to be revised within 5 years.
Footnotes
Acknowledgement: The authors declare no conflict of interest. This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Funding: Ruben Pauwels is supported by the European Union Horizon 2020 Research and Innovation Programme under the Marie Skłodowska-Curie grant agreement number 754513 and by Aarhus University Research Foundation (AIAS-COFUND).
Contributor Information
Hugo Gaêta-Araujo, Email: hugogaeta@hotmail.com.
Tamara Alzoubi, Email: Alzo3bi.tam@outlook.com.
Karla de Faria Vasconcelos, Email: karlafav13@gmail.com.
Kaan Orhan, Email: call53@yahoo.com.
Ruben Pauwels, Email: pauwelsruben@hotmail.com.
Jan W Casselman, Email: Jan.Casselman@azsintjan.be.
Reinhilde Jacobs, Email: reinhilde.jacobs@kuleuven.be.
REFERENCES
- 1.Nemtoi A, Czink C, Haba D, Gahleitner A. Cone beam CT: a current overview of devices. Dentomaxillofac Radiol 2013; 42: 20120443Available from. doi: 10.1259/dmfr.20120443 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Jain S, Choudhary K, Nagi R, Shukla S, Kaur N, Grover D. New evolution of cone-beam computed tomography in dentistry: combining digital technologies. Imaging Sci Dent . 2019; 49: 179Available from[Internet]. doi: 10.5624/isd.2019.49.3.179 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Kiljunen T, Kaasalainen T, Suomalainen A, Kortesniemi M. Dental cone beam CT: a review. Physica Medica 2015; 31: 844–60Available from. doi: 10.1016/j.ejmp.2015.09.004 [DOI] [PubMed] [Google Scholar]
- 4.Hayashi T, Arai Y, Chikui T, Hayashi-Sakai S, Honda K, Indo H, et al. . Clinical guidelines for dental cone-beam computed tomography. Oral Radiol . 2018; 34: 89–104Available from[Internet]. doi: 10.1007/s11282-018-0314-3 [DOI] [PubMed] [Google Scholar]
- 5.Scarfe W. C, Angelopoulos C. Maxillofacial Cone Beam Computed Tomography [Internet]. Cham: Springer International Publishing; 2018. Available from: http://link.springer.com/10.1007/978-3-319-62061-9. [Google Scholar]
- 6.Lamoral Y, Quirynen M, Peene P, Vanneste P, Lemahieu S, Baert A, et al. . Computed tomography in the preoperative planning of oral endo-osseous implant surgery. RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren 1990; 153: 505–9Available from. doi: 10.1055/s-2008-1033428 [DOI] [PubMed] [Google Scholar]
- 7.Quirynen M, Lamoral Y, Dekeyser C, Peene P, van Steenberghe D, Bonte J, et al. . Ct scan standard reconstruction technique for reliable jaw bone volume determination. Int J Oral Maxillofac Implants . 1990; 5: 384–9Available from[Internet]. [PubMed] [Google Scholar]
- 8.Garg AK, Vicari A. Radiographic modalities for diagnosis and treatment planning in implant dentistry. Implant Soc . 1995; 5: 7–11Available from[Internet]. [PubMed] [Google Scholar]
- 9.Mozzo P, Procacci C, Tacconi A, Tinazzi Martini P, Bergamo Andreis IA. A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol . 1998; 8: 1558–64Available from[Internet]. doi: 10.1007/s003300050586 [DOI] [PubMed] [Google Scholar]
- 10.Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of a compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol 1999; 28: 245–8Available from. doi: 10.1038/sj.dmfr.4600448 [DOI] [PubMed] [Google Scholar]
- 11.Horner K, Jacobs R, Schulze R. Dental CBCT equipment and performance issues. Radiat Prot Dosimetry . 2013; 153: 212–8Available from[Internet]. doi: 10.1093/rpd/ncs289 [DOI] [PubMed] [Google Scholar]
- 12.Scarfe WC, Li Z, Aboelmaaty W, Scott SA, Farman AG. Maxillofacial cone beam computed tomography: essence, elements and steps to interpretation. Aust Dent J . 2012; 57(Suppl 1): 46–60Available from[Internet]. doi: 10.1111/j.1834-7819.2011.01657.x [DOI] [PubMed] [Google Scholar]
- 13.Daly MJ, Siewerdsen JH, Moseley DJ, Jaffray DA, Irish JC. Intraoperative cone-beam CT for guidance of head and neck surgery: assessment of dose and image quality using a C-arm prototype. Med Phys . 2006; 33: 3767–80Available from[Internet]. doi: 10.1118/1.2349687 [DOI] [PubMed] [Google Scholar]
- 14.Scarfe WC, Farman AG. What is cone-beam CT and how does it work? Dent Clin North Am . 2008; 52: 707–30Available from[Internet]. doi: 10.1016/j.cden.2008.05.005 [DOI] [PubMed] [Google Scholar]
- 15.Korreman S, Rasch C, McNair H, Verellen D, Oelfke U, Maingon P, et al. . The European Society of therapeutic radiology and Oncology–European Institute of radiotherapy (ESTRO–EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide. Radiotherapy and Oncology . 2010; 94: 129–44Available from[Internet]. doi: 10.1016/j.radonc.2010.01.004 [DOI] [PubMed] [Google Scholar]
- 16.Rehani MM, Gupta R, Bartling S, Sharp GC, Pauwels R, Berris T, et al. . ICRP publication 129: radiological protection in cone beam computed tomography (CBCT. Ann ICRP . 2015; 44: 7–127Available from[Internet]. doi: 10.1177/0146645315575485 [DOI] [PubMed] [Google Scholar]
- 17.Pauwels R, Araki K, Siewerdsen JH, Thongvigitmanee SS. Technical aspects of dental CBCT: state of the art. Dentomaxillofac Radiol 2015; 44: 20140224Available from. doi: 10.1259/dmfr.20140224 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Kau CH, BozÌŒicÌŒ M, English J, Lee R, Bussa H, Ellis RK. Cone-Beam computed tomography of the maxillofacial region—an update. Int. J. Med. Robotics Comput. Assist. Surg. 2009; 5: 366–80Available from. doi: 10.1002/rcs.279 [DOI] [PubMed] [Google Scholar]
- 19.Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, et al. . Effective dose range for dental cone beam computed tomography scanners. Eur J Radiol . 2012; 81: 267–71Available from[Internet]. doi: 10.1016/j.ejrad.2010.11.028 [DOI] [PubMed] [Google Scholar]
- 20.De Vos W, Casselman J, Swennen GRJ. Cone-Beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: a systematic review of the literature. Int J Oral Maxillofac Surg . 2009; 38: 609–25Available from[Internet]. doi: 10.1016/j.ijom.2009.02.028 [DOI] [PubMed] [Google Scholar]
- 21.Sedentexct Radiation Protection 172: Cone Beam CT for Dental and Maxillofacial Radiology - Evidence-based Guidelines. Off Off Publ Eur Communities [Internet] 2012; 156. [Google Scholar]
- 22.Ludlow JB, Timothy R, Walker C, Hunter R, Benavides E, Samuelson DB, et al. . Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol 2015; 44: 20140197Available from. doi: 10.1259/dmfr.20140197 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Bornstein MM, Scarfe WC, Vaughn VM, Jacobs R. Cone beam computed tomography in implant dentistry: a systematic review focusing on guidelines, indications, and radiation dose risks. Int J Oral Maxillofac Implants 2014; 29 Suppl(Supplement): 55–77. doi: 10.11607/jomi.2014suppl.g1.4 [DOI] [PubMed] [Google Scholar]
- 24.Yajima A, Otonari-Yamamoto M, Sano T, Hayakawa Y, Otonari T, Tanabe K, et al. . Cone-Beam CT (CB Throne) applied to Dentomaxillofacial region. Bull Tokyo Dent Coll . 2006; 47: 133–41Available from[Internet]. doi: 10.2209/tdcpublication.47.133 [DOI] [PubMed] [Google Scholar]
- 25.Nasseh I, Al-Rawi W. Cone beam computed tomography. Dent Clin North Am . 2018; 62: 361–91Available from[Internet]. doi: 10.1016/j.cden.2018.03.002 [DOI] [PubMed] [Google Scholar]
- 26.Goulston R, Davies J, Horner K, Murphy F. Dose optimization by altering the operating potential and tube current exposure time product in dental cone beam CT: a systematic review. Dentomaxillofac Radiol 2016; 45: 20150254Available from. doi: 10.1259/dmfr.20150254 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Oenning AC, Jacobs R, Pauwels R, Stratis A, Hedesiu M, Salmon B. Cone-Beam CT in paediatric dentistry: DIMITRA project position statement. Pediatr Radiol . 2018; 48: 308–16Available from[Internet]. doi: 10.1007/s00247-017-4012-9 [DOI] [PubMed] [Google Scholar]
- 28.The 2007 recommendations of the International Commission on radiological protection. ICRP publication 103. Ann ICRP 2007; 37(2-4): 1-332. doi: 10.1016/j.icrp.2007.10.003 [DOI] [PubMed] [Google Scholar]
- 29.Pauwels R, Jacobs R, Bogaerts R, Bosmans H, Panmekiate S. Reduction of scatter-induced image noise in cone beam computed tomography: effect of field of view size and position. Oral Surg Oral Med Oral Pathol Oral Radiol . 2016; 121: 188–95Available from[Internet]. doi: 10.1016/j.oooo.2015.10.017 [DOI] [PubMed] [Google Scholar]
- 30.Pauwels R, Beinsberger J, Stamatakis H, Tsiklakis K, Walker A, Bosmans H, et al. . Comparison of spatial and contrast resolution for cone-beam computed tomography scanners. Oral Surg Oral Med Oral Pathol Oral Radiol . 2012; 114: 127–35Available from[Internet]. doi: 10.1016/j.oooo.2012.01.020 [DOI] [PubMed] [Google Scholar]
- 31.Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM. Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health . 2018; 18: 88Available from[Internet]. doi: 10.1186/s12903-018-0523-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Dawood A, Patel S, Brown J. Cone beam CT in dental practice. Br Dent J . 2009; 207: 23–8Available from[Internet]. doi: 10.1038/sj.bdj.2009.560 [DOI] [PubMed] [Google Scholar]
- 33.Spin-Neto R, Wenzel A. Patient movement and motion artefacts in cone beam computed tomography of the dentomaxillofacial region: a systematic literature review. Oral Surg Oral Med Oral Pathol Oral Radiol . 2016; 121: 425–33Available from[Internet]. doi: 10.1016/j.oooo.2015.11.019 [DOI] [PubMed] [Google Scholar]
- 34.Jacobs R. Dental cone beam CT and its justified use in oral health care. J Belgian Soc Radiol [Internet] 2011; 94: 254Available from. doi: 10.5334/jbr-btr.662 [DOI] [PubMed] [Google Scholar]