Table 3. Contrast-to-noise (CNR) and figure of merit (FOM) values of different nodules at the threshold of scanning conditions.

Kilovolt Tolerance threshold of subjective perception to image quality Image quality threshold of −800 HU nodules observed
Milliampere-second Volume CT dose index (mGy) CNR FOM Milliampere-second Volume CT dose index (mGy) CNR FOM
A B C D A B C D A B C D A B C D
120 20 1.4 2.7 1.2 1.4 0.9 5.0 1.0 1.4 0.6 10 0.7 1.9 0.9 1.2 0.6 5.2 1.2 2.3 0.6
100 40 1.6 3.1 1.6 1.7 1.0 6.1 1.7 1.8 0.6 20 0.8 2.5 1.0 1.5 0.6 7.9 1.1 2.7 0.5
80 80 1.6 3.0 1.4 1.0 0.8 5.7 1.3 0.6 0.4 40 0.8 2.4 1.1 1.2 0.5 7.5 1.4 1.8 0.4

A, −630 HU nodule with 12-mm diameter at right chest wall of right lower lung; B, −800 HU nodule with 8-mm diameter at posterior chest wall of right lower lung, adjacent to spine; C, −630 HU nodule with 8-mm diameter at left costophrenic angle; D, −800 HU nodule with 12-mm diameter at left costophrenic angle.