ICP-OES instrument parameters for determination of Cd and Pb.

Metal ion Power (kW) Plasma gas flow (L min−1) Auxiliary gas flow (L min−1) Nebulizer gas flow rate (L min−1) Delay time (sec) Wavelength (nm) Detection limit (μg L−1)
Cd 1.4 15 1.5 0.7 12 226.502 0.2
Pb 220.353 1.8