Skip to main content
. 2020 Nov;8(21):1454. doi: 10.21037/atm-20-7047

Table 1. In-batch precision and total precision of Mindray’s fully automated chemiluminescent immunoassay system.

Sample Mean X¯ (ng/mL) Within-day precision SD Within-batch precision Total precision
SD CV (%) SD CV (%)
Level 1 0.50 0.013 0.010 2.07 0.015 3.05
Level 2 9.42 0.158 0.079 0.83 0.171 1.81

SD, standard deviation; CV, coefficient of variation.