Table 5. Results of the comparison of the different systems.

PCT test results segmented diagnostic reference system compliance rate
Concentration (ng/mL) Roche (case) CL900I-1 (case) Snibe 2000 CL900I compliance rate Snibe 2000 compliance rate
<0.1 5 4 0 80.00% 0.00%
0.1–0.25 5 5 2 100.00% 40.00%
0.25–0.5 7 7 1 100.00% 14.29%
>0.5 23 23 23 100.00% 100.00%