. 2020 Sep 23;68(10):2122–2125. doi: 10.4103/ijo.IJO_546_19

Table 2.

Comparison of Agreement of IOP Using the Icare ic100, Icare TA01i, and Goldmann Applanation Tonometer (GAT)

Group ICC 95% CI 95% LoA Cronbach’s alpha
Icare ic100 vs. 7-16 0.872 (0.786-0.923) −3.69, 3.03 0.874
Icare TA01i >16-<23 0.785 (0.600-0.885) −4.37, 3.18 0.797
≥23 0.988 (0.978-0.993) −2.79, 3.21 0.988
Overall 0.989 (0.985-0.992) −3.64, 3.18 0.989
Icare TA01i vs. GAT 7-16 0.818 (0.695-0.891) −3.39, 3.72 0.816
>16-<23 0.713 (0.471-0.845) −3.27, 4.13 0.718
≥23 0.976 (0.948-0.988) −2.95, 4.74 0.979
Overall 0.987 (0.981-0.990) −4.11, 3.43 0.987
GAT vs. Icare ic100 7-16 0.876 (0.793-0.926) −3.08, 3.42 0.876
>16-<23 0.528 (0.138-0.744) −3.18, 5.23 0.574
≥23 0.979 (0.960-0.989) −2.95, 4.33 0.981
Overall 0.987 (0.980-0.991) −3.13, 4.28 0.988