Skip to main content
. 2020 Sep 23;68(10):2122–2125. doi: 10.4103/ijo.IJO_546_19

Table 3.

Repeatability of IOP measurement using Icare ic100, Icare TA01i, and Goldmann Applanation Tonometer (GAT)

Icare ic100 Icare TA01i GAT
ICC 0.997 0.998 0.998
95% CI Lower 0.997 0.997 0.997
95% CI Upper 0.998 0.998 0.998
Sw 0.96 0.88 0.92
95% CI Lower 0.85 0.78 0.82
95% CI Upper 1.07 0.98 1.03
RC 2.67 2.43 2.55
95% CI Lower 2.37 2.15 2.26
95% CI Upper 2.97 2.70 2.84
CVw (%) 4.89 4.36 4.54
95% CI Lower 4.33 3.86 4.02
95% CI Upper 5.44 4.85 5.05

Sw is within subject standard deviation and CVw is the within-subject coefficients of variation