Table 2. Quantitative fit testing results of N95 masks after repeat decontamination cycles.

PortaCount Result (normal & deep breathing exercises only)
Groups Masks
Control 3M 1860 pass
3M Aura 1870 pass
3M Vflex 1804S pass
AO Safety 1054S pass
3M 8210 pass
3M 9210 pass
# of cycles
1 3 5 10
Autoclave 3M 1860 pass fail fail fail
3M Aura 1870 pass pass pass pass
3M Vflex 1804S pass pass pass pass
AO Safety 1054S pass pass pass pass
3M 8210 pass fail fail fail
3M 9210 pass pass pass pass
1 3
EtO 3M 1860 pass pass
3M Aura 1870 pass pass
3M Vflex 1804S pass pass
AO Safety 1054S pass pass
1 2 5 10
LT-HPGP 3M 1860 pass fail fail fail
3M Aura 1870 pass fail fail fail
3M Vflex 1804S pass fail fail fail
AO Safety 1054S pass pass fail fail
3M 8210 pass fail fail fail
3M 9210 pass fail fail fail
1 3 5 10
VHP 3M 1860 pass pass pass pass
3M Aura 1870 pass pass pass pass
3M Vflex 1804S pass pass pass pass
AO Safety 1054S pass pass pass pass
1 3 5 10
PAF 3M 1860 pass pass pass pass
3M Aura 1870 pass pass pass pass
3M Vflex 1804S pass pass pass pass
AO Safety 1054S pass pass pass pass
1 3 5 10
UV-C (1120 mJ/cm2) 3M 1860 pass pass pass ND
3M Aura 1870 pass pass pass ND
3M Vflex 1804S pass pass pass ND
AO Safety 1054S pass pass pass ND
3M 8210 pass pass pass ND
3M 9210 pass pass pass ND
1 3 5 10
Moist Heat (75°C & 22% RH X 3 hr) 3M 1860 ND pass ND pass
3M Aura 1870 ND pass ND pass
3M Vflex 1804S ND pass ND pass
AO Safety 1054S ND pass ND pass
3M 8210 ND pass ND pass
3M 9210 ND pass ND pass

PAF: Peracetic acid dry fogging system, VHP = vaporous hydrogen peroxide, EtO = ethylene oxide.

LT-HPGP: Low temperature hydrogen peroxide gas plasma.

UV-C: Ultraviolet light-C radiation (254 nm wavelength).