Skip to main content
. 2020 Nov 18;11(24):4447–4459. doi: 10.1021/acschemneuro.0c00648

Table 1. Limits of Detection and Intraday and Interday Precision Values Obtained for Analyzed Elements.

parameter P S K Ca Fe Cu Zn Se
LOD (SD) [ppm] 12.96(0.45) 8.32(0.29) 4.41(0.16) 1.832(0.072) 0.1916(0.0082) 0.0852(0.0035) 0.1078(0.0042) 0.0297(0.0011)
intraday precision [%] 2.0 1.6 1.0 3.1 0.3 0.7 0.2 4.9
interday precision [%] 1.3 1.2 0.7 2.8 0.3 0.4 0.2 1.2