Table 12.

Propagation of Raman photons from source to detector for gypsum ν1 emission

Distance (m) At source At aperture At cathode
1.8 1.8e7 4820 310
2.3 4.0e7 4080 260
2.8 7.5e7 3480 220
3.3 1.0e8 2400 150
3.8 1.3e8 1710 110
4.3 1.3e8 1080 68
4.8 1.3e8 690 43
6.0 1.1e8 250 14

Units are photons between the FWHM of the Raman peak per laser pulse. For the source emission, only photons produced within the spectrometer FOV are counted (as in Fig. 49), resulting in an increase between 1.8 and 3.8 m. Numbers are based on measurements made with the EQM. The FM will have different results based on alignment at the time of measurement