. 2020 Dec 21;217(1):4. doi: 10.1007/s11214-020-00777-5

Table 3.

Optical characteristics of the spectrometers and their detectors

Ultraviolet (UV) Violet (VIO) Transmission
Techniques LIBS LIBS, VIS LIBS, Raman, TRL, VIS
Spectrometer type - - - Crossed Czerny-Turner - - - Transmission
Spectral range 243.5–341.7 nm 382.1–467.5 nm 535–620 nm, green
620–712 nm, orange
712–853 nm, red
Pixel resolution 0.0478 nm/pixel 0.0417 nm/pixel 0.064–0.090 nm/pixel
Channels 2048 2048 2048 × 3
Vertical integration rows 200: 151–351 200: 151–351 Red: 91–161; 21–221a
40: 232–272 40: 236–276 Green: 236–356
16: 252–268 16: 248–264 Orange: 378–506
CCD coating Enhanced UV Enhanced Broadband Enhanced Midband
CCD QEb 77.2% @ 250 nm 81.3% @ 345 nm 92.8% @ 635 nm
64.9% @ 300 nm
55.4% @ 345 nm
CCD dark current (e-/pix/sec @ 300 K) 269 202 224
Photon response non-uniformity (PRNU) 4.5% @ 350 nm 0.7% @ 400 nm 0.7% @ 650 nm
Full well, single pixel 123k e 128k e 128k e
Full well serial reg. 190k e 192k e 192k e
Conversion gain (e-/DNc) 3.14 ± 0.06 3.16 ± 0.05 3.15 ± 0.06

aQE, dark currents, and PRNU were measured by the CCD manufacturer

bRed region has options of 70 or 200 row integrations

cDigital numbers, also known as analog-to-digital units (ADU), or counts