Table 7.

Transmission spectrometer CCD operation steps

Action 1/rate (μs) Number Duration (ms) Total (ms)
Integrationa 5000.00 5.00 5.00
Dump unexposed rowsb 17.99 92 1.65 6.65
Transfer red band to serial register 17.99 70 1.26 7.91
Digitize serial register 2.50 2148 5.37 13.28
Dump unexposed rows 17.99 74 1.20 14.61
Transfer green band to serial register 17.99 120 2.16 16.77
Digitize serial register 2.50 2148 5.37 22.14
Dump unexposed rows 17.99 22 0.40 21.54
Transfer orange band to serial register 17.99 128 2.30 24.84
Digitize serial register 2.50 2148 5.37 30.21

aDoes not include intensifier exposure duration, which should be added on for long gates

bCalibration has been carried out for two settings of the red band. The one shown here, with 70 rows, cuts off some of the light and provides higher resolution