Skip to main content
. 2021 Mar 1;93(5):2838–2847. doi: 10.1002/jmv.26696

Table 1.

Performance of test kit on national reference panels

Reference panel Positive compliance rate Negative compliance rate Limit of detection Repeatability
IgM antibody test National reference panel Criterion 10/10+/+ 24/25−/− ≥2/10+/+, L1+, L2+ 10/10+/+
Results 10/10+/+ 24/25−/− 3/10+/+, L1+, L2+, L3+ 10/10+/+
IgG antibody test National reference panel Criterion ≥9/10+/+ 24/25−/− ≥1/10+/+, L1+ 10/10+/+
Results 9/10+/+ 25/25−/− 1/10+/+, L1+ 10/10+/+