Table 3:

Relative errors in relaxation times for different acquisition patterns when T1=1400ms and T2=50ms, which are values representative of myocardial tissue at 3T. Each row refers to a physical effect or system imperfection that was modeled when generating the ground truth cMRF signal timecourses but not modeled when matching to the dictionary.

Dictionary Corrections cMRF Acquisition Pattern
A B C D E F G H I J

Neglecting slice profile and preparation pulse efficiency; Ground truth B1+ is 100% T1 Error (%) −5.5 −8.3 −8.3 −8.3 −8.3 −10.5 −8.3 −8.3 −5.5 −11.1
T2 Error (%) −4.2 5.2 −4.2 0.0 10.4 15.6 0.0 0.0 −6.2 62.5

Neglecting preparation pulse efficiency; Ground truth B1+ is 100% T1 Error (%) 0.0 −2.4 0.0 0.0 0.0 0.0 0.0 0.0 −0.5 0.0
T2 Error (%) 5.2 4.5 5.2 5.2 0.0 0.0 5.2 5.2 6.1 0.0

Neglecting B1+; Ground truth B1+ is 70% T1 Error (%) 2.8 0.0 0.0 0.0 0.0 0.0 0.0 0.0 2.2 −2.8
T2 Error (%) −4.2 1.0 −4.2 0.0 5.2 10.4 −3.6 0.0 −4.2 46.9