Example CLM1B Blue LED Calibration Curve

(A) Calibration curve for 24-well LAVA board with two different light guide thicknesses.

(B) Calibration curve for a 96-well LAVA board. Light intensity recorded using a PM100D compact power and energy meter console (Thorlabs) with a standard photodiode power sensor (S120C, Thorlabs). Curve slope and y-intercept used in software design to designate individual well intensity.