Table 5.

Number of Patient Exclusions Based on Diameter, Length, and Angle Measurements of the Aortic Arch and Supra‐Aortic Vessels

Aortic Diameters, mm*

Mean±SD or

Median (Range)

Exclusions, N (%)

Relay Plus Double‐Branch Device

(Terumo)

TAG Thoracic Branch Endo‐Prosthesis

(Gore)

Zenith Arch Branched Device

(Cook)

Nexus Stent Graft System (Endospan/

Jotec)

Proximal LZ (zero point) (n=59) 36.1 (25.6–74.9) 11 (19)
Proximal LZ (zone 0+45 mm) (n=18) 35.4±3.8 0 (0)
Proximal LZ (zone 0+60 mm) (n=38) 42.3 (30.7–81.3) 12 (32)
Proximal LZ (zone 0–total) (n=76) 38.4 (26.0–81.3) 10 (18) 41 (54)
Proximal LZ (zone 0–BCA) (n=153) 44.5±11.4 100 (65)
Proximal LZ (zone 1) (n=23) 34.2±3.7 0 (0)
Proximal LZ (zone 2) (n=22) 32.7±4.1 0 (0)
Proximal BCA (n=76) 17.5 (12.7–30.3) 9 (15) 21 (37) 8 (11)
Distal BCA+25 mm (n=57)§ 14.1 (10.8–21.1) 1 (2)
Distal BCA (n=153) 18.8±4.6 67 (43)
Distal LCCA (n=75)|| 7.5±1.2 19 (33) 0 (0)
Maximal LSA (n=22) 13.4±2.1 1 (1)
Aortic lengths, mm
STJ to zero point (n=76) 96.3±22.5 3 (5) 0 (0)|| 17 (22)
Zero point to distal LCCA (n=59) 38.1±6.2 10 (17)
Mid BCA to distal LCCA (n=23) 20.0±3.9 18 (78)
Distal BCA to distal LCCA (n=23) 12.4±3.7 20 (87)
Mid LCCA to distal LSA (n=22) 28.3±4.8 3 (14)
Distal LCCA to distal LSA (n=22) 23.4±4.7 3 (14)
BCA ostium to bifurcation (n=76) 33.5 (14.5–62.9) 2 (3) 5 (9)||

LSA to first branch of LSA (n=22)

STJ to BCA (n=153)

38.7±10.1

102.6±27.4

3 (14)

0 (0)

Distal BCA LZ length (n=153) 34.7±12.8 11 (7)
Branch clock positions/angles (degrees)
BCA clock position (n=76) 3.8 (−43.3 to 19.1) 1 (0) 7 (9)
LCCA clock position (n=76) 13.5±14.5 1 (0) 2 (3)
LSA clock position (n=76) 7.1±16.7
BCA angle to aortic arch (n=165) 141.0±23.3 30 (18)

BCA indicates brachiocephalic artery; LCCA, left common carotid artery; LSA, left subclavian artery; LZ, landing zone; and STJ, sinotubular junction.

a

Measurements composed of 110 TAAs, 253 aortic dissections, of which 12 were chronic dissections and the rest acute, and 14 other pathological features.

b

A total of 56 of 59 measured; 3 patients were ineligible for either 45‐ or 60‐mm proximal length device because of insufficient BCA–zero point length.

c

A total of 23 of 30 measured; 7 patients were excluded from zone 1 proximal LZ length measurements because of proximity of BCA.

d

A total of 57 of 59 measured; 2 patients did not have sufficient BCA length to measure 25 mm distally.

e

A total of 75 of 76 (58/59 Relay Branch eligible) patients measured; 1 patient had a stenosis of LCCA.

f

A total of 57 of 76 patients were subjected to zone 0 assessment with the TAG Thoracic Branch Endoprosthesis (Gore).