Skip to main content
. 2020 Dec 12;25(24):5885. doi: 10.3390/molecules25245885

Table 4.

Parameter setting for pre-processing of LC-MS data using MZmine software (VTT Technical Research Centre, Finland).

Parameters Details
Baseline correction Mass Spectrometry (MS) level = 1
m/z bin width = 1
Asymmetric baseline corrector (smoothing = 100,000, asymmetry = 0.5)
Mass detection Mass detector = centroid (noise level = 200)
Chromatogram builder Min time span = 0.2 min
Min height = 200
m/z tolerance = 1.0 mz or 2500 ppm
Peak detection Filter width = 11
Isotopic peaks grouper m/z tolerance = 1.0 mz or 2500 ppm
Retention time (RT) tolerance = 0.2 min
Monotonic shape
Maximum charge = 1
Duplicate peak filter m/z tolerance = 1.0 mz or 2500 ppm
RT tolerance = 0.2 min
Normalization Linear Normalization
Peak measurement type = peak area
Alignment (Join aligner) m/z tolerance = 1.0 or 2500 ppm
Weight for m/z = 85
RT tolerance = 0.2 min
Weight for RT = 15
Require same charge state
Gap filling Same RT and m/z range gap filler
m/z tolerance = 1.0 or 2500 ppm