Skip to main content
. 2020 Dec 23;12:293–299. doi: 10.2147/DHPS.S280006

Table 1.

Malfunctions, Injuries, and Deaths Related to Use of Elastomeric Pumps to Deliver Local Anesthetics, Including Reports of LAST and Injuries of Interest

Total Reports Involving Elastomeric Pumps to Deliver Local Anesthetic (n=4093)
Reports identifying pumps with flow issues 903 (22.1)
Reports identifying possible or probable LAST 139 (3.4)
Reports identifying possible liver injury 13 (0.3)
Reports identifying possible surgical site infection 51 (1.2)
Malfunction reports 3624 (88.5)
 Possible or probable LAST 26 (0.6)
 Possible liver injury 2 (<0.1)
 Possible site infection 5 (0.1)
Injuries 292 (7.1)
 Possible or probable LAST 84 (2.1)
 Possible liver injury 11 (0.3)
 Possible site infection 38 (0.9)
Deaths 8 (0.2)
 Possible or probable LAST 2 (<0.1)
 Possible liver injury 0 (0.0)
 Possible site infection 0 (0.0)
Other report codes/missing codes 169 (4.1)
 Possible or probable LAST 27 (0.7)
 Possible liver injury 0 (0.0)
 Possible site infection 8 (0.2)

Note: Values are the n (%).

Abbreviation: LAST, local anesthetic systemic toxicity.