Table 7.
Effects of Different Instruments, Columns, Column Temperatures, Flow Rates, Volume and pH on RRTs (n = 3)
| Effects Factor | RPZ | PPZ | IPZ | OPZ(EOPZ) | LPZ | |
|---|---|---|---|---|---|---|
| RSD% | Chromatographic Columns | 2.70 | 3.96 | 2.75 | – | 4.75 |
| HPLC Instruments | 2.65 | 1.52 | 1.34 | – | 0.89 | |
| Column Temperatures | 2.35 | 1.20 | 0.60 | – | 0.68 | |
| Flow Volume of Mobile Phase | 1.14 | 1.07 | 1.01 | – | 2.04 | |
| Flow Rate of Mobile Phase | 0.27 | 0.24 | 0.48 | – | 0.44 | |
| pH Value of Mobile Phase | 15.48 | 2.34 | 13.72 | – | 3.07 | |
| χ | 0.69 | 1.29 | 1.79 | – | 1.98 | |
| S | 0.003 | 0.005 | 0.007 | – | 0.008 | |
| χ±3S | 0.69±0.009 | 1.29±0.015 | 1.79±0.021 | – | 1.98±0.024 |