Skip to main content
. 2020 Nov 12;21(12):188–196. doi: 10.1002/acm2.13086

Table 3.

Summary of the mean ± SD rmsTRE of the image registration methods applied at cranial and body sites.

rmsTRE
Cranial (N = 20) Body (N = 16)
PMTB 0.4 ± 0.1 mm 0.7 ± 0.1 mm
PMEX 0.6 ± 0.1 mm 0.7 ± 0.1 mm
3D3DV 0.5 ± 0.1 mm 1.2 ± 0.2 mm
3D3DE 0.6 ± 0.2 mm 1.1 ± 0.3 mm
2D3D 0.9 ± 0.2 mm 1.5 ± 0.1 mm
3DS 1.2 ± 0.2 mm 1.0 ± 0.1 mm