Skip to main content
. 2020 Oct 30;21(12):62–73. doi: 10.1002/acm2.13060

Table 1.

Mean absorbed doses and the 68% confidence interval (CI) for the target breast from WBRT alone and WBRT plus RNRT for full MC PHITS and TPS.

Mean dose for target breast

Gamma passing rate

(3%/3 mm) [%]

MC [Gy] TPS [Gy] Δ [%]
Left side irradiation WBRT 52.100 ± 0.594 51.666 −0.833 100
WBRT plus RNRT 51.275 ± 0.728 51.579 0.593 100
Right side irradiation WBRT 51.258 ± 0.569 51.462 0.398 100
WBRT plus RNRT 51.101 ± 0.572 51.589 0.956 99.8

Difference Δ is given by the equation [Δ = (PHITS ‐ TPS)/ TPS × 100]. Gamma passing rates were calculated between PHITS and TPS.