. 2020 Jul 17;15(1):E11–E16. doi: 10.5489/cuaj.6712

Table 3.

Detailed results per site

Site Multiparametric MRI Micro-ultrasound n


True positive True negative False positive False negative True positive True negative False positive False negative
 A 70 46 64 14 84 13 97 0 194
 B 3 1 5 0 3 0 6 0 9
 C 97 22 163 10 97 44 141 10 292
 D 9 7 1 2 10 2 6 1 19
 E 76 5 76 2 74 12 69 4 159
 F 20 9 20 2 22 5 24 0 51
 G 24 13 38 2 25 9 42 1 77
 H 29 27 53 7 31 24 56 5 116
 I 19 3 25 0 17 8 20 2 47
 J 21 1 40 0 19 21 20 2 62
 K 3 2 8 1 4 0 10 0 14
Total 371 136 493 40 386 138 491 25 1040

While clear site-level variability is seen on targeting percentage and accuracy, only 3 sites (B, I, J) failed to achieve non-inferior sensitivity on their own. Aggregate results demonstrate superior sensitivity for micro-ultrasound (p=0.03) and non-inferior specificity (p<0.01). MRI: magnetic resonance imaging.