Skip to main content
. 2020 Dec 17;59(1):e01262-20. doi: 10.1128/JCM.01262-20

TABLE 3.

Discrepant analysis of 4 samples

Sample identifier SAMBAb
PHE Colindale assay RdRp Initial (CT) PHE Cambridge (Wuhan) assay (UTM sample)
PHE Cambridge (Wuhan) assay (SCoV sample)
Final result
Initial result Repeat result Initial RdRp (CT) Repeat RdRp (CT) E (CT) Initial RdRp (CT) Repeat RdRp (CT) E (CT)
1-25 NEG NEG POS (31.18) POS (28.87) POS (28.96) Not tested Not tested Not tested Not tested POS
2-17 POS POS NEG NEG NEG POS (34.99) POS (33.98) NEG NEG POS
2-25 POS POS NEG (36.58)a POS (33.62) NEG NEG NEG NEG NEG POS
2-38 POS POS NEG NEG POS (35.09) POS (34.06) NEG NEG NEG POS
a

This specimen was assessed as negative by the Colindale RdRp assay; however, when the assay run’s raw data were reexamined and the normal assay control parameters were relaxed (i.e., threshold and endpoint fluorescence) it was weakly positive (shallow curve), with a threshold cycle (CT) value of 36.58.

b

POS, positive; NEG, negative.