TABLE 2.

Distribution of oxygen saturation and measured difference by instrument

Measurement Instrument
Measure Control OxyWatch SM Onyx
Total patients (N) 200 198 200 200
Oxygen saturation (%)
Mean (SD)* 96.08% (3.84) 94.23% (4.70) 95.02% (4.49) 95.08% (3.90)
Min, Max SpO2 59%, 100% 59%, 100% 62%, 100% 68%, 100%
Pearson Correlation Coefficient (P value) Ref 0.79 (< 0.001) 0.76 (< 0.001) 0.79 (0.001)
Difference in SpO2 (Control‐Measured; %)
Mean (95% CI) Ref 1.87% (1.46–2.28) 1.07% (0.65–1.48) 1.00% (0.65–1.35)
Min, Max Ref −5%, 19% −6%, 23% −9%, 15%

CI, confidence interval; SM, Santa Medical.

*Analysis of variance: P = 0 < 0.001; F value = 15.85.