TABLE 3.

Distribution of oxygen saturation and measured differences by instrument in patients with control SpO2 ≤ 94% (N = 60)

Measurement Instrument
Measure Control OxyWatch SM Onyx
Measured SpO2
Mean (SD) a 92.15% (4.57) 90.25% (5.76) 91.37% (5.10) 91.60% (4.68)
Min, Max SpO2 59%, 94% 59%, 99% 62%, 98% 68%, 100%
Pearson's Correlation Coefficient (P value) Ref 0.75 (0.001) 0.79 (0.001) 0.70 (0.001)
Difference in measured SpO2 (Control‐Measured)
Mean (95% CI) Ref 1.95% (0.96–2.94) 0.78% (−0.04–1.60) 0.55% (−0.37–1.47)
Min, Max Ref −5%, 13% −6%, 10% −9%, 15%

CI, confidence interval; SM, Santa Medical.

a

Analysis of variance: P < 0.0001, F value = 14.49.