Table 3. Results of 8-sample pooled testing, by testing platform and number of positive specimens per pool (n = 110) for detection of severe acute respiratory syndrome coronavirus 2*.

Pool no.
Pooled testing
Individual testing
Total no. pools
LDT
Panther Fusion
Panther Aptima-M
Panther Aptima-350
Positive, 
(no. 1 PP, no. >1 PP)
Negative
1 + + + + 30 (21, 9) 0 30
2 + + + 2 (1, 1) 0 2
3 + + 0 (0, 0) 0 0
4 + + + 0 (0, 0) 0 0
5 + + 0 (0, 0) 0 0
6 + 1 (1, 0) 1† 2
7 + + + 2 (2, 0) 0 2
8 + + 1 (1, 0) 0 1
9 + 0 (0, 0) 1‡ 1
10 + + 2 (2, 0) 0 2
11 + 1 (1, 0) 2
12 7 (7, 0) 61 68
No. positive pools 34 36 34 39 46 (36, 10)
No. negative pools 76 74 76 71 64
Total no. pools 110 110 110 110 110

*Panther Aptima with manufacturer-set relative light unit cutoff. Panther Aptima-350, Panther Aptima with relative light unit cutoff value >350 was considered positive; Ct, cycle threshold; LDT, laboratory-developed test; Pos, positive; RLU, relative light unit; 1 PP, 1 positive specimen in pool; >1 PP, >2 positive specimens in pool; –, negative; +, positive.

†False-positive LDT Ct value was 37.5.

‡False-positive Panther Fusion Ct value was 38.8.

§False-positive Panther Aptima-350 RLU value was 434.