Skip to main content
. 2020 Dec 31;15(12):e0244590. doi: 10.1371/journal.pone.0244590

Table 1. Mean value, mean difference and 95% limit of agreement of axial length in 175 eyes and anterior chamber depth in 104 eyes readings of the two biometry devices.

Mean value ± SD (mm) Mean difference ± SD (95% confidence interval) (mm) p-value* 95% LoA
Axial length PCI 24.45 ± 0.14 -0.03 ± 0.09 (-0.04, -0.02) <0.001 0.36 (-0.15 to 0.21)
SS-OCT 24.42 ± 0.13
ACD PCI 3.44 ± 0.54 0.06 ±0.10 (0.04, 0.08) <0.001 0.40 (-0.14 to 0.26)
SS-OCT 3.50 ± 0.57

*: p-value of paired T-test between two devices.

SD: standard deviation; LoA: limits of agreement; PCI: partial coherence interferometry, SS-OCT: swept-source optical coherence tomography; ACD: anterior chamber depth.