Skip to main content
. 2021 Jan 6;8:1. doi: 10.1186/s40658-020-00344-5

Table 2.

General patient characteristics and main PET parameters recorded or computed in the different groups imaged in clinical routine with the Vereos (digital-PET) or Ingenuity (analog-PET) camera and after the injection of 3 MBq.kg-1 of current 18F-labeled tracers, with an additional patient imaged after the injection of 7 MBq.kg-1 of 82Rb on the Vereos camera. Continuous variables are presented with median values [minimum – maximum values after exclusion of outliers]

Ingenuity TF Vereos
Tracer type 18F-FDG 18F-Choline 18F-DOPA 18F-FDG 18F-Choline 18F-DOPA 82Rb
Number of patients 10 10 1 10 10 1 1
Gender Female Male Female Female Male Female Male
Age (years)

76

[67-82]

70

[67-75]

72

71

[71-77]

72

[70-76]

65 55
Body weight (kg)

68

[65-69]

78

[74-84]

84

63

[59-69]

80

[77-81]

90 159
Injection-to-recording delay-time (min)

69

[65-72]

14

[12-17]

0

67

[65-74]

24

[17-25]

0 0
Recorded area Whole body

Whole

body

Pelvis Whole body

Whole

body

Pelvis Thorax
Number of list-mode samples 18534 20757 2382 24814 21210 2401 961

Single count rate

(Mcps)

8.1

[5.8-10]

14.2

[7.4-21.7]

19.2

[17.7-20.1]

4.6

[2.3-6.0]

7.6

[3.7-12.6]

10.5

[7.9-13.3]

4.6

[0.5 29.5]

Corresponding activity concentration on NEMA phantom (kBq.mL-1)

2.9

[1.7-3.9]

6.2

[2.5-10.4]

9.0

[8.1-9.5]

2.9

[1.3-3.9]

5.1

[2.3-8.7]

7.2

[5.3-9.2]

3.0

[0.3-20.8]

Corresponding TOF resolution (ps)

631

[621-643]

677

[628-732]

713

[702-719]

328

[325-332]

Corresponding energy resolution (%)

11.36

[11.28-11.43]

11.6

[11.34-11.82]

11.7

[11.68-11.77]

11.01

[10.98-11.05]