. 2021 Jan 6;20:4. doi: 10.1186/s12938-020-00842-8

Table 2.

Results of the RFA for different values of applied voltage on a 2-cm spherical zone in case of non-doped tissue (ND) and two types of dopants: saline solution (0.9% NaCl) and AuNPs (0.01 wt%)

Dopant Zi (Ω) 50 V 70 V 90 V
t100-Ω B A t100-Ω B A t100-Ω B A
ND 87.7 457 2.7 4.4 94.5 1.6 3.8 31.5 1.0 3.7
NaCl
 1% 84.6 454 2.8 4.3 95 1.7 4.0 35 1.2 3.7
 2% 82 452 2.8 4.3 97.5 1.7 4.0 35.5 1.2 3.7
 5% 75.5 457 3.1 4.6 103 1.9 4.0 36 1.3 3.8
 10% 67.8 450 3.5 5.0 109.5 2.2 4.2 37.5 1.3 3.9
AuNPs
 1% 78 456 3.0 4.5 101 1.8 4.0 35.5 1.2 3.6
 2% 71.5 451 3.2 4.7 106.5 2.0 4.1 36.5 1.3 3.8
 5% 60 448 3.7 4.8 120.5 2.6 4.2 41 1.5 4.0
 10% 51.6 479 4.4 5.0 144 3.2 4.4 49.5 2.3 4.0

Zi: initial impedance; t100-Ω: time until roll-off (in s); B and A: transverse and axial diameters (in cm) of the coagulation zone, respectively. The different values of concentration (from 1 to 10% are volume fraction)