Table 4.

Results of the RFA for different values of applied voltage on a 4-cm spherical zone in case of non-doped tissue (ND) and two types of dopants: saline solution (0.9% NaCl) and AuNPs (0.01 wt%)

Dopant Zi (Ω) 50 V 70 V 90 V
t100-Ω B A t100-Ω B A t100-Ω B A
ND 87.7 440 2.7 4.4 94.5 1.6 3.8 31.5 1.0 3.7
NaCl
1% 81.4 400 2.7 4.4 87 1.6 3.9 30 1.2 3.6
2% 76 367 2.6 4.4 80 1.7 3.9 29 1.2 3.6
5% 64 306 2.5 4.3 66 1.7 3.9 25.5 1.2 3.6
10% 52 269 2.5 4.4 55 1.7 3.8 22.5 1.2 3.4
AuNPs
1% 69 327 2.5 4.4 71 1.7 3.9 27 1.2 3.6
2% 58 287 2.5 4.4 60 1.7 3.9 23.5 1.2 3.6
5% 42 245 2.5 4.4 50.5 1.6 3.8 20.5 1.2 3.6
10% 31.9 265 2.6 4.9 54.5 1.7 3.9 21.5 1.2 3.6

Zi: initial impedance; t100-Ω: time until roll-off (in s); B and A: transverse and axial diameters (in cm) of the coagulation zone, respectively. The different values of concentration (from 1 to 10% are volume fraction)