Table 2.
Figure number | Sample | Matrix and protocol | Rotor type | Year | Build-up time (s) | MAS (kHz) | εon/off |
---|---|---|---|---|---|---|---|
Figure 2a | WT | 13C-depleted, d8-glycerol/D2O/H2O (60/30/10 vol%) | ZrO2 thin-wall rotor | 2018 | 2.77 | 10 | 26 |
Additional file 1: Fig.S2a | ctl1 ctl2 | 13C-depleted, d8-glycerol/D2O/H2O (60/30/10 vol%) | 2.59 | 10 | 22 | ||
Figure 3a | ctl1 | 13C-depleted, d8-glycerol/D2O/H2O (60/30/10 vol%) | Sapphire rotor | 2020 | 2.64 | 8 | 38 |
Figure 3a | ctl1 | d6-DMSO/D2O/H2O (80/10/10 vol%) | 1.96 | 8 | 48 | ||
Figure 3a | ctl1 | D2O/H2O(75/25 vol%) | – | 8 | 42 | ||
Figure 3a | ctl1 | D2O (matrix-free) | 1.67 | 8 | 57 | ||
Additional file 1: Fig. S2b | ctl1 ctl2 | D2O (matrix-free) | 1.40 | 10.5 | 44 |
The related figure position of each matrix protocol is given. Unidentified (–). The samples freshly measured in 2020 give high sensitivity enhancement of 38–57 fold due to the use of sapphire rotor, optimization of sample preparation, and the improvement of the DNP instruments