Table 2.

Electrode measurements across all patients.

Patient number Electrode number Electrode location Position Electrode location confirmation Radial target error Depth error Entry point radial error Trajectory length
1 1 L HC Supine CT 2.7 0.0 0.5 92.6
2 1 R HC Supine CT 1.8 3.7 1.8 86.0
2 L HC 2.4 3.3 0.4 88.9
3 1 L HC Supine CT 0.8 4.4 0.5 102.9
4 1 L HC Supine O-arm 3.0 5.8 0.1 97.4
2 L para HC 2.7 −1.7 0.0 93.0
5 1 R HC Supine CT 0.4 −0.9 1.0 102.2
2 L HC 2.1 −6.3 1.1 103.5
8 1 R HC Supine CT 1.0 −0.8 1.9 98.5
2 L HC 2.0 −3.2 1.8 99.1
6 1 R HC Prone O-arm 0.7 −2 0.5 83.3
2 L HC 0.9 −2.9 2.1 83.5
7 1 R HC Prone O-arm 0.3 5.3 0.2 85.7
2 L HC 0.6 −1.2 0.9 87.7
9 1 R HC Prone O-arm 0.6 −1.8 0.5 89.8
2 L HC 1.1 −7 0.0 92.9
10 1 R HC Prone O-arm 1.2 −0.4 0.8 83.1
2 L HC 0.7 −1 1.5 81.9

All measurements in mm. Electrodes placed in the prone position had smaller radial target error compared to those placed in the supine position. HC, hippocampus.