Table 1.

Detailed Characteristics of the Four Refractive Error Measurement Techniques Evaluated in this Study

Characteristics Retinoscope Tabletop Autorefractor Photoscreener E-see
Working principle Foucault knife-edge Image size assessment using rotary prisms Eccentric IR photorefraction Wavefront aberrometry
Operating range No specific operating range +22 D to -25 D for sphere and ±10.00 D for cylinder ±7.50 D for sphere and ±3.50 D for cylinder ±10.00 D for sphere and ±6.00 D for cylinder
Invasiveness of measurement Noninvasive. Measurements performed from a distance. Noninvasive. Participant stabilized on forehead rest. Noninvasive. Measurements are performed from a distance. Device touches the face for measurement
Portability Portable Non-portable Portable Portable
Measurement time Consumes time for measurement ∼5 measurements per second <1 s per measurement ∼10 measurements per second
Fixation distance Distant target Near target with simulated distance viewing Distant target Distant target
Pupil size dependence Measurements become challenging with pupil miosis Stable measurements for pupil diameter up to 2 mm Measurement accuracy decreases with pupil miosis Measurement accuracy may vary with pupil diameter
Binocular viewing / measurements Viewing can be binocular, but the measurement is monocular Only monocular viewing and measurements Viewing and measurement can be binocular Only monocular viewing and measurements
Participant cooperation Needs limited cooperation. Useful in challenging cases. Participant cooperation is essential for reliable measurements Needs limited cooperation. Useful in challenging cases. Participant cooperation is critical for reliable measurements
Examiner training Examiner training is time-consuming Examiner can be trained within a short time Examiner can be trained within a short period of time Examiner can be trained within a short period of time
Near-triad measurements Only measurement of accommodation is possible Only measurement of accommodation is possible Near-triad can be measured in sync with each other Only measurement of accommodation is possible

Data on operating ranges and measurement time are obtained from the manufacturer prescribed user manual of each instrument.