Skip to main content
. 2020 Dec 24;13(1):31. doi: 10.3390/cancers13010031

Table 2.

Minimum performance metrics for successful MRT in hyperthermia treatments.

Measure Metric Definition Minimum
Bias Mean error (ME) ME=1nj=1n(EjA) ≤|0.5 °C|
Spatial temperature precision Spatial temperature standard deviation (SD) SD2=1nj=1n(EjE¯)2 ≤0.5 °C
Temporal temperature precision Temporal temperature standard deviation (SD) Variability at different time points over 90 min ≤0.5 °C
Accuracy Mean absolute error (MAE) MAE=1nj=1n|EjA| ≤1 °C