Table 3.

Fiber-optic cantilevered scanners developed for endomicroscopy.

Actuation Principle Imaging Modality Scanning Direction Resolution FOV Frequency Driving Voltage Frame Rate Scanner Dimensions Scanning Pattern Cantilever Fiber Refs.
Piezoelectric bimorph Multiphoton Forward 0.8 µm (lateral), 10 µm (axial) 110 µm × 110 µm 1.05 kHz (fast axis), 4.1 Hz 50 Vpp (resonant bimorph) 200 Vpp (non-resonant) 4.1 fps (512 × 512 pixels) 3 mm (diameter)
40 mm (rigid length)
Raster scan DCF [105]
Piezoelectric tube Two-photon Forward 0.61 µm ÷ 1.10 µm (from center of FOV to peripheric zone) 160 µm 3.1 kHz 48 Vpp 6 fps ~2 mm (diameter) Spiral DCF with GRIN lens [156]
Piezoelectric tube Two-Photon Forward 10 µm 70° 5 kHz 25 V 15 fps 1.6 mm (diameter) Spiral SMF [16]
Piezoelectric thin film E-OCT Side-view 5 µm (axial) 152° 394 Hz 1.3 Vpp - 3.4 mm × 2.5 mm radial - [157]
Electrothermal Confocal endomicroscope Forward ~1.7 µm 378 µm × 439 µm 239 Hz (x-axis) 207 Hz (y-axis) 16 Vpp 1 fps 1.65 mm (diameter) 28 mm (rigid length) Lissajous SMF [65]
Electrothermal OCT (A-scan) Forward ~17 µm (lateral)
~9 µm (axial)
~3 mm (beam scanning) ~100 Hz 3 Vac_pp, 1.5 V DC offset 200 fps 5.5 mm (diameter) 55 mm (rigid length) raster - [96]
Electromagnetic OCT (B-scan) Forward 4–6 µm (axial) 25–35 µm (lateral) 2 mm 5 Hz ±10 V (triangle wave) - 0.51 mm (diameter) Linear SMF [132]
Micromotor PA and US endoscopy Side-view ~58 µm (PA radial) ~30 µm (US radial)
~100 µm (PA transvers) ~120 µm (US transverse)
~310° ~4 Hz ~3.2V DC 4 fps 2.5 mm (diameter)
~35 mm (Rigid length)
Radial MMF [40]
Electromagnetic Confocal Forward 0.8 µm (lateral) 390 µm × 390 µm 700 Hz (fast scan) 1–2 Hz (slow scan) - 1 fps 8 mm (diameter) Raster SMF [158]
Cellvizio Confocal Forward 5–15 µm (axial) 2–5 µm (lateral) 600 µm × 500 µm - - 12 fps 2.5 mm - - [33]