Skip to main content
. 2020 Dec 30;10(1):108. doi: 10.3390/jcm10010108

Table 1.

Accuracy and precision for the analysis of p-Coumaric acid in human plasma.

Nominal (ng/mL) Precision CV (%) Accuracy (%)
Intra (n = 5) Inter (n = 3) Intra (n = 5) Inter (n = 3)
0.2 5.6 6.4 103.8 108.4
0.6 2.3 2.4 100.0 100.9
8 1.2 2.1 99.0 99.6
16 1.0 1.3 99.2 101.3