. 2021 Jan 7;58(3):174–180. doi: 10.1177/0004563220984827

Table 2.

Accuracy of results from rapid antibody tests regarding concordance with quantitative antibody test results.

ALLTest SD BIOSENSOR KURABO
(n  =  69) (n  =  55) (n  =  34)
IgG antibody
 Concordance rate (%) 76.8 k = 0.55 96.4 k = 0.93 88.2 k = 0.72
 Sensitivity (%) 67.4 χ2 = 23.38 96.7 χ2 = 47.23 72.7 χ2 = 17.88
 Specificity (%) 92.3 P < 0.01 96 P < 0.01 95.7 P < 0.01
 Positive predictive value (%) 93.5   96.7   88.9  
 Negative predictive value (%) 64.2   96   88  
 Threshold IgG titre (AU/mL) 19.1   19.1   19.1  
 Test positive rate over IgG threshold titre 69   100   100  
IgM antibody
 Concordance rate (%) 55.1 k = 0.07 88.7 k = 0.76 91.2 k = 0.62
 Sensitivity (%) 15.2 χ2 = 0.78 85 χ2 = 31.34 75 χ2 = 13.14
 Specificity (%) 91.7 P  =  0.47 90.9 P < 0.01 93.3 P < 0.01
 Positive predictive value (%) 62.5   85   60  
 Negative predictive value (%) 54.1   90.9   96.6  
 Threshold IgM titre (AU/mL) 13.8   12.2   12.2  
 Test positive rate over threshold titre 17.9   89.5   75